Comparison of peripapillary retinal nerve fiber layer thickness measured by spectral vs. time domain optical coherence tomography.

Hong Samin, Seong Gong Je, Kim Sung Soo, Kang Sung Yong, Kim Chan Yun

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the peripapillary retinal nerve fiber layer (RNFL) thickness measured by spectral and time domain optical coherence tomography (OCT).

PATIENTS AND

METHODS

One hundred and eighty-eight subjects (normal tension glaucoma, n = 80; primary open angle glaucoma, n = 49; healthy controls, n = 59) were included in this study. Peripapillary RNFL thickness was measured by an Optic Disc Cube 200 × 200 scan of the spectral domain Cirrus HD OCT and by a Fast RNFL scan of the time domain Stratus OCT. They were compared using linear regression (R(2)), Bland-Altman plots, weighted Kappa (K(w)) coefficients, and areas under the receiver operating characteristic curve (AROC).

RESULTS

Except for the inferior quadrant, RNFL thickness measurements by Cirrus HD OCT were thinner than Stratus OCT (all p < 0.001); for the inferior quadrant, this tendency was vice versa (p  =  0.043). Average RNFL thicknesses of the two OCTs showed a good correlation (R(2), 0.852) but not excellent agreement (mean of difference, 4.0 μm; +1.96 SD, 10.7 μm; -1.96 SD, -18.6 μm). Regarding the probability codes, only the average RNFL thickness had almost perfect agreement (K(w), 0.983). In addition, Cirrus HD OCT had the largest AROC for the superior quadrant (0.795), but Stratus OCT had the largest AROC for average RNFL thickness (0.764) (p  =  0.281).

CONCLUSIONS

Except for the inferior quadrant, the average and quadrants RNFL thickness measured by the Cirrus HD OCT were thinner than the Stratus OCT. Regarding discriminating power for glaucoma detection, the Cirrus HD OCT was better than the Stratus OCT for the superior and inferior quadrant RNFL thicknesses but not for the average RNFL thickness.

Для цитирования: Hong Samin, Seong Gong Je, Kim Sung Soo, Kang Sung Yong, Kim Chan Yun. Comparison of peripapillary retinal nerve fiber layer thickness measured by spectral vs. time domain optical coherence tomography.. Current eye research. 2010. DOI: 10.3109/02713683.2010.533807.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.