DOI 10.1097/ijg.0b013e3181e66498

Медицинская библиотека · Journal of glaucoma · 2011

Goldmann applanation tonometry and dynamic contour tonometry are not correlated with central corneal thickness in primary open angle glaucoma.

Regev Guy, Harris Alon, Siesky Brent, Shoshani Yochai, Egan Patrick, Moss Adam, Zalish Miriam, WuDunn Darrell, Ehrlich Rita

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare intraocular pressure (IOP) measured by Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT) and assess their relationship to central corneal thickness (CCT) in patients with primary open angle glaucoma (OAG).

PATIENTS AND

METHODS

GAT, DCT, and CCT were assessed in 116 patients with OAG [mean age 65.9 (10.5); 59% female] participating in the Indianapolis Glaucoma Progression Study. GAT and DCT were measured in a randomized order followed by CCT (ultrasonic corneal pachymetry) during a single study visit. Bland-Altman plots were used to evaluate the limits of agreement between tonometery methodologies whereas multivariate regression analysis was used to evaluate the influence of CCT on GAT and DCT IOP measurements.

RESULTS

IOP values obtained by DCT and GAT showed a strong positive correlation in patients with OAG (r=0.93; P<0.001). Mean IOP measured with DCT [18.4 (5.1) mm Hg] was significantly higher (P<0.001) than GAT IOP measurements [16.5 (4.5) mmHg]. CCT did not seem to influence either GAT or DCT measurements (r=0.1025, P=0.16; r=0.05, P=0.46), respectively. The Bland-Altman data showed that the amount of disagreement between IOP assessment techniques varied, suggesting a proportional bias.

CONCLUSIONS

In this group of patients with OAG, there was a strong correlation between GAT and DCT measurements of IOP. IOP measured with DCT was consistently higher than IOP measured with GAT. Neither GAT nor DCT measurements were correlated with CCT. This data suggests that factors other than CCT may be involved in the tendency of DCT to produce higher measures of IOP than GAT.

Для цитирования: Regev Guy, Harris Alon, Siesky Brent, Shoshani Yochai, Egan Patrick, Moss Adam, Zalish Miriam, WuDunn Darrell, Ehrlich Rita. Goldmann applanation tonometry and dynamic contour tonometry are not correlated with central corneal thickness in primary open angle glaucoma.. Journal of glaucoma. 2011. DOI: 10.1097/ijg.0b013e3181e66498.