European Society for Neurogastroenterology and Motility (ESNM) recommendations for the use of high-resolution manometry of the esophagus.
Аннотация
BACKGROUND
Several patients in gastroenterology practice present with esophageal symptoms, and in case of normal endoscopy with biopsies, high-resolution manometry (HRM) is often the next step. Our aim was to develop a European consensus on the clinical application of esophageal HRM, to offer the clinician guidance in selecting patients for HRM and using its results to optimize clinical outcome.
METHODS
A Delphi consensus was initiated with 38 multidisciplinary experts from 16 European countries who conducted a literature summary and voting process on 71 statements. Quality of evidence was evaluated using grading of recommendations assessment, development, and evaluation (GRADE) criteria. Consensus (defined as >80% agreement) was reached for 33 statements.
RESULTS
The process generated guidance on when to consider esophageal HRM, how to perform it, and how to generate the report. The Delphi process also identified several areas of uncertainty, such as the choice of catheters, the duration of fasting and the position in which HRM is performed, but recommended to perform at least 10 5-ml swallows in supine position for each study. Postprandial combined HRM impedance is considered useful for diagnosing rumination. There is a large lack of consensus on treatment implications of HRM findings, which is probably the single area requiring future targeted research.
CONCLUSIONS AND INFERENCES: A multinational and multidisciplinary group of European experts summarized the current state of consensus on technical aspects, indications, performance, analysis, diagnosis, and therapeutic implications of esophageal HRM.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.