Bioprosthetic Valve Thrombosis Versus Structural Failure: Clinical and Echocardiographic Predictors.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Egbe Alexander C, Pislaru Sorin V, Pellikka Patricia A, Poterucha Joseph T, Schaff Hartzell V, Maleszewski Joseph J, Connolly Heidi M
Аннотация
BACKGROUND: Bioprosthetic valve thrombosis (BPVT) is considered uncommon; this may be related to the fact that it is often unrecognized. Recent data suggest that BPVT responds to vitamin K antagonists, emphasizing the need for reliable diagnosis.
OBJECTIVES: This study sought to determine the diagnostic features of BPVT and to formulate a diagnostic model for BPVT.
METHODS: Cases of BPVT occurring between 1997 and 2013 were identified from the Mayo Clinic pathology database. Patients with BPVT were matched 1:2 for age, sex, and prosthesis position with patients whose valves were explanted for structural failure. We formulated a diagnostic model for BPVT using multivariate linear logistic regression and receiver operating characteristic.
RESULTS: Among 397 consecutive cases of explanted bioprostheses, there were 46 cases of BPVT (11.6%; aortic 29, mitral 9, tricuspid 7, pulmonary 1), mean age was 63 years, and 68% were male. Thirty (65%) cases occurred >12 months post-implantation; median bioprosthetic valve longevity was 24 months (cases) versus 108 months (controls) (p < 0.001). Independent predictors of BPVT were >50% increase in mean echo-Doppler gradient from baseline within 5 years (odds ratio [OR]: 12.7), paroxysmal atrial fibrillation (OR: 5.19), subtherapeutic international normalized ratio (OR: 7.37), increased cusp thickness (OR: 12.2), and abnormal cusp mobility (OR: 6.94). Presence of all 5 diagnostic features was predictive of BPVT with 76% sensitivity, 93% specificity, 85% positive predictive value, and 89% negative predictive value (p < 0.001).
CONCLUSIONS: BPVT is not uncommon and can occur several years after surgery. A combination of clinical and echocardiographic features can reliably diagnose BPVT.
OBJECTIVES: This study sought to determine the diagnostic features of BPVT and to formulate a diagnostic model for BPVT.
METHODS: Cases of BPVT occurring between 1997 and 2013 were identified from the Mayo Clinic pathology database. Patients with BPVT were matched 1:2 for age, sex, and prosthesis position with patients whose valves were explanted for structural failure. We formulated a diagnostic model for BPVT using multivariate linear logistic regression and receiver operating characteristic.
RESULTS: Among 397 consecutive cases of explanted bioprostheses, there were 46 cases of BPVT (11.6%; aortic 29, mitral 9, tricuspid 7, pulmonary 1), mean age was 63 years, and 68% were male. Thirty (65%) cases occurred >12 months post-implantation; median bioprosthetic valve longevity was 24 months (cases) versus 108 months (controls) (p < 0.001). Independent predictors of BPVT were >50% increase in mean echo-Doppler gradient from baseline within 5 years (odds ratio [OR]: 12.7), paroxysmal atrial fibrillation (OR: 5.19), subtherapeutic international normalized ratio (OR: 7.37), increased cusp thickness (OR: 12.2), and abnormal cusp mobility (OR: 6.94). Presence of all 5 diagnostic features was predictive of BPVT with 76% sensitivity, 93% specificity, 85% positive predictive value, and 89% negative predictive value (p < 0.001).
CONCLUSIONS: BPVT is not uncommon and can occur several years after surgery. A combination of clinical and echocardiographic features can reliably diagnose BPVT.