Spinal cord · 2003
Subdural catheter migration may lead to baclofen pump dysfunction.
Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Pasquier Y, Cahana A, Schnider A
Аннотация
OBJECTIVES: To report an unusual cause of intrathecal drug delivery failure in baclofen pump device.
STUDY DESIGN: A case report of an SCI patient treated with intrathecal baclofen, presenting a drug withdrawal.
SETTING: Regional spinal cord injuries centre in Geneva (Switzerland).
METHODS: We present a case of a 38-year-old male with complete T9 spastic paraplegia for 15 years, treated with intrathecal baclofen for 11 years. He recently presented to our centre with a spastic hypertonic episode, associated with rhabdomyolysis.
RESULTS: Standard investigations were unrevealing. However, a CT scan performed after injecting a radio-opaque solution by the side port of the pump, showed an unexpected catheter migration into the subdural space. Surgical revision reversed withdrawal symptoms.
CONCLUSIONS: Subdural catheter migration must be considered in the differential diagnosis of intrathecal drug delivery system failures. We recommend the use of the CT scan after contrast injection, to detect the localization of the distal catheter tip and confirm the normal diffusion into the subarachnoid space.
STUDY DESIGN: A case report of an SCI patient treated with intrathecal baclofen, presenting a drug withdrawal.
SETTING: Regional spinal cord injuries centre in Geneva (Switzerland).
METHODS: We present a case of a 38-year-old male with complete T9 spastic paraplegia for 15 years, treated with intrathecal baclofen for 11 years. He recently presented to our centre with a spastic hypertonic episode, associated with rhabdomyolysis.
RESULTS: Standard investigations were unrevealing. However, a CT scan performed after injecting a radio-opaque solution by the side port of the pump, showed an unexpected catheter migration into the subdural space. Surgical revision reversed withdrawal symptoms.
CONCLUSIONS: Subdural catheter migration must be considered in the differential diagnosis of intrathecal drug delivery system failures. We recommend the use of the CT scan after contrast injection, to detect the localization of the distal catheter tip and confirm the normal diffusion into the subarachnoid space.
Предметные рубрики MeSH
Accidental FallsAdultBaclofenCatheterizationEquipment FailureFollow-Up StudiesForeign-Body MigrationHumansInfusion Pumps, ImplantableMaleMuscle Relaxants, CentralMuscle SpasticityParaplegiaRehabilitation CentersReoperationRisk AssessmentSpinal Cord InjuriesSubstance Withdrawal SyndromeThoracic VertebraeTomography, X-Ray ComputedTreatment Outcome