Intrasubject repeatability of corneal power, thickness, and wavefront aberrations with a new version of a dual rotating Scheimpflug-Placido system.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Cerviño Alejandro, Dominguez-Vicent Alberto, Ferrer-Blasco Teresa, García-Lázaro Santiago, Albarrán-Diego Cesar

Аннотация

PURPOSE: To determine the intrasubject repeatability of a recently introduced dual-camera rotating Scheimpflug-Placido imaging system (Galilei G4) in determining corneal thickness, power, and wavefront aberrations in young healthy subjects.

SETTING: University of Valencia, Valencia, Spain.

DESIGN: Prospective evaluation of diagnostic technology.

METHODS: The study comprised right eyes of 25 subjects aged 20 to 40 years with a spherical equivalent ranging from -4.25 to +1.00 diopters. The central corneal thickness, thinnest point value and location, anterior and posterior surface curvatures, total corneal power, and corneal wavefront aberrations were measured for distance vision using the dual Scheimpflug-Placido system. Three consecutive measurements were taken in each eye.

RESULTS: Twenty-five eyes were evaluated. Repeated-measures analysis of variance showed the only statistically significant difference between the 3 repeated measures to be in trefoil aberration. Intraclass correlation coefficients (ICCs) were higher than 0.950 for all the parameters except the thinnest point chord distance to geometric corneal center (0.528) and chord angle (0.742), corneal astigmatism (0.811) and its vector components J0 (0.891) and J45 (0.724), and all wavefront aberrations.

CONCLUSIONS: The new dual Scheimpflug-Placido system had high intraobserver repeatability for corneal power and thickness and moderate repeatability for corneal astigmatism and corneal wavefront aberrations in healthy corneas with low astigmatism. This iteration of the device performed better in young healthy corneas than preceding versions.

FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.

Предметные рубрики MeSH