Long-term clinical outcomes of toric intraocular lens implantation in cataract cases with preexisting astigmatism.
СтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Miyake Toshiyuki, Kamiya Kazutaka, Amano Rie, Iida Yoshihiko, Tsunehiro Shuntaro, Shimizu Kimiya
Аннотация
PURPOSE: To evaluate the long-term clinical outcomes and rotational stability of toric intraocular lenses (IOLs) to correct preexisting astigmatism in cataract patients.
SETTING: Kitasato University Hospital, Kanagawa, Japan.
DESIGN: Prospective observational study.
METHODS: Phacoemulsification with implantation of a toric IOL (Acrysof IQ toric SN6AT) was performed in cataract patients with corneal astigmatism. The postoperative follow-up was up to 2 years.
RESULTS: This study enrolled 378 eyes of 302 patients with a mean age of 63.4 years ± 16.9 (SD). The corrected distance visual acuity 3 months postoperatively was 20/25 or better in 94.7% of eyes. The mean refractive cylinder was -1.92 ± 1.45 diopters (D) preoperatively, -0.59 ± 0.62 D 3 months postoperatively (322 eyes), and -0.67 ± 0.90 D at 2 years (73 eyes). There were statistically significant differences between the preoperative and 3-month postoperative measurements (P<.001, Wilcoxon signed-rank test). The mean IOL misalignment was 4.1 ± 3.0 degrees 2 years postoperatively. The mean IOL rotation was 4.5 ± 4.9 degrees within 1 day postoperatively. The rotation was more than 20 degrees in 6 eyes, all of which had an axial length (AL) of more than 25.0 mm. All rotations occurred within 10 days postoperatively.
CONCLUSIONS: Toric IOLs were effective in reducing preexisting corneal astigmatism and had overall good rotational stability. A large degree of IOL rotation might occur in eyes with a relatively long AL, especially during the early postoperative period. The 6 rotated IOLs were implanted to correct with-the-rule astigmatism.
FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
SETTING: Kitasato University Hospital, Kanagawa, Japan.
DESIGN: Prospective observational study.
METHODS: Phacoemulsification with implantation of a toric IOL (Acrysof IQ toric SN6AT) was performed in cataract patients with corneal astigmatism. The postoperative follow-up was up to 2 years.
RESULTS: This study enrolled 378 eyes of 302 patients with a mean age of 63.4 years ± 16.9 (SD). The corrected distance visual acuity 3 months postoperatively was 20/25 or better in 94.7% of eyes. The mean refractive cylinder was -1.92 ± 1.45 diopters (D) preoperatively, -0.59 ± 0.62 D 3 months postoperatively (322 eyes), and -0.67 ± 0.90 D at 2 years (73 eyes). There were statistically significant differences between the preoperative and 3-month postoperative measurements (P<.001, Wilcoxon signed-rank test). The mean IOL misalignment was 4.1 ± 3.0 degrees 2 years postoperatively. The mean IOL rotation was 4.5 ± 4.9 degrees within 1 day postoperatively. The rotation was more than 20 degrees in 6 eyes, all of which had an axial length (AL) of more than 25.0 mm. All rotations occurred within 10 days postoperatively.
CONCLUSIONS: Toric IOLs were effective in reducing preexisting corneal astigmatism and had overall good rotational stability. A large degree of IOL rotation might occur in eyes with a relatively long AL, especially during the early postoperative period. The 6 rotated IOLs were implanted to correct with-the-rule astigmatism.
FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.