Comparison of anterior segment measurements with optical low-coherence reflectometry and rotating dual Scheimpflug analysis.

Huerva Valentín, Ascaso Francisco J, Soldevila Jordi, Lavilla Laura

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the anterior segment measurements obtained with optical low-coherence reflectometry (OLCR) (Lenstar LS 900, version 2.1.1) and rotating dual Scheimpflug analysis (Galilei G1, version 3).

SETTING

Private practice, Policlinic Lleida, Lleida, Spain.

DESIGN

Prospective comparative observational study.

METHODS

The following measurements were performed in patients requiring a preoperative study for a refractive procedure or cataract surgery and healthy volunteers from the clinic's staff: central corneal thickness (CCT), anterior chamber depth (ACD), horizontal limbal distance (white-to-white distance [WTW]), pupil diameter, keratometry (K) readings at the steepest meridian (steep K) and the flattest meridian (flat K), corneal astigmatism power, and plus astigmatic cylinder. Measurements were taken using the OLCR device and rotating dual Scheimpflug analyzer. The main outcome measure was the degree of agreement between steep K, flat K, astigmatism power, cylinder axis, CCT, ACD, WTW, and pupil diameter measurements.

RESULTS

The study enrolled 100 eyes of 100 subjects. Steep K, flat K, corneal astigmatism power, cylinder axis, and WTW measurements with the OLCR device and Scheimpflug analyzer showed narrow 95% limits of agreement (LoA), which implies good agreement (P>.05, Bland-Altman plot analysis). In contrast, the range and 95% LoA for CCT, ACD, and pupil diameter values were statistically significantly different (P<.05, Bland-Altman plot analysis).

CONCLUSION

In clinical practice, the OLCR device and the rotating dual Scheimpflug analyzer system can be used interchangeably for WTW measurements and K readings but not for CCT, ACD, and pupil diameter values.

FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.

Для цитирования: Huerva Valentín, Ascaso Francisco J, Soldevila Jordi, Lavilla Laura. Comparison of anterior segment measurements with optical low-coherence reflectometry and rotating dual Scheimpflug analysis.. Journal of cataract and refractive surgery. 2014. DOI: 10.1016/j.jcrs.2013.10.045.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.