One-year refractive results, contrast sensitivity, high-order aberrations and complications after myopic small-incision lenticule extraction (ReLEx SMILE).
Аннотация
BACKGROUND
To report one year results of the first cohort of routine refractive lenticule extraction through a small incision (ReLEx SMILE) for correction of myopia and myopic astigmatism.
METHODS
Fifty-four eyes of 27 patients with spherical equivalent of -4.68 ± 1.29D who underwent routine ReLEx SMILE by a single surgeon were prospectively followed-up for 1 year. We used the VisuMax femtosecond laser system (Carl Zeiss Meditec AG, Germany) with a 500 kHz repetition rate. Folow-up intervals were at 1 day, 1 week, 1, 3, 6, and 12 months after surgery. We obtained following parameters: uncorrected (UDVA) and distance-corrected visual acuity (CDVA), contrast sensitivity, and wave front measurements. We also recorded all complications.
RESULTS
Because of suction loss in one eye, 12-month results were obtained in 53 eyes as follows. After 1 year, 88% of eyes with plano target had an UDVA of 20/20 or better. Twelve percent of eyes lost 1 line of CDVA, while 31% gained 1 line and 3% gained 2 lines. The mean SE after 1 year was -0.19 ± 0.19. The mean refraction change between month 1 and 12 was 0.08 D. Neither mesopic nor photopic contrast sensitivity showed any significant changes. The high-order aberrations (HOA) increased from 0.17 to 0.27 μm (Malacara notation). No visually threatening complications were observed.
CONCLUSION
In this first cohort, ReLEx SMILE produced satisfactory refractive outcomes with moderate induction of HOA and unaffected contrast sensitivity after 1 year.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.