Inspiratory effort and measurement of dynamic intrinsic PEEP in COPD patients: effects of ventilator triggering systems.

Ranieri V M, Mascia L, Petruzzelli V, Bruno F, Brienza A, Giuliani R

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

To investigate effects of ventilator triggering systems (pressure and flow triggering: PT and FT) on measurement of dynamic intrinsic PEEP (PEEPidyn) and patient-ventilator interaction in patients with chronic obstructive pulmonary disease during weaning from mechanical ventilation.

DESIGN

Prospective study.

SETTING

Medical/surgical intensive care unit of an academic hospital.

PATIENTS AND

PARTICIPANTS

6 COPD patients with acute respiratory failure ready to wean.

MEASUREMENTS: We measured flow, airway opening, esophageal and gastric pressures. Minute ventilation, breathing pattern and pressure time product (PTP) of the respiratory muscles and of the diaphragm were obtained during spontaneous ventilation through a mechanical ventilator (Puritan-Bennett 7200ae). Two triggering systems, namely PT and FT, were evaluated.

RESULTS

The inspiratory muscles effort necessary to overcome the triggering system overestimated PEEPidyn measurement of an amount equal to 49 +/- 2 and 58 +/- 3% during respectively pressure and flow triggering. FT increased tidal volume and minute ventilation and decrease PTP/b and PTP/min of the respiratory muscles and diaphragm.

CONCLUSIONS

To correctly measure PEEPidyn, the inspiratory effort produced to overcome PEEPi and to trigger the ventilator must be discriminated. Application of flow triggering requires less effort to initiate inspiration and provide a positive end-expiratory pressure level that is able to unload the respiratory muscles by reducing PEEPi. With flow triggering higher minute ventilation are obtained in COPD patients during the weaning phase.

Для цитирования: Ranieri V M, Mascia L, Petruzzelli V, Bruno F, Brienza A, Giuliani R. Inspiratory effort and measurement of dynamic intrinsic PEEP in COPD patients: effects of ventilator triggering systems.. Intensive care medicine. 1995. DOI: 10.1007/bf01712330.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.