PURPOSE: To investigate the intraoperative performance and lens fragmentation efficacy of a non-cavitating handheld lensectomy system in mild, moderate, and severe cataract.

SETTING

Ambulatory surgical centers.

DESIGN

Retrospective consecutive case series.

METHODS

665 consecutive eyes underwent cataract surgery by 12 surgeons using a new handheld non-cavitating lensectomy system for nuclear fragmentations and extraction. Intraoperative measurements included surgical time, miLOOP pretreatment, and irrigation fluid use.

RESULTS

Of the 665 eyes, 38 (6%), 468 (70%), 126 (19%), and 33 (5%) were of grade 1, 2, 3, and 4 nuclear densities, respectively, as graded by the surgeon intraoperatively. Successful nuclear fragmentation, lens extraction, and cortical removal were achieved in all eyes. Total nucleus fragmentation and extraction times were 70.1 seconds, 100.3 seconds, 132.6 seconds, and 287.9 seconds for grades 1, 2, 3, and 4, respectively ( P < .001). In addition, irrigation and aspiration cortical removal times were 64.1 seconds, 51.1 seconds, 48.5 seconds, and 59.0 seconds, respectively ( P = .14). There was a low rate of capsular tear (3 cases in 665 surgeries, 0.45%) and no other emergent adverse events.

CONCLUSIONS

The miCOR handheld non-cavitating lensectomy system demonstrated nuclear fragmentation and extraction in the absence of intraocular cavitation across all grades of nuclear densities.

Для цитирования: Ianchulev Tsontcho, Yeu Elizabeth, Hu Edward H, Singh Paul, Kamthan Gautam, Li Gavin, Pantanelli Seth, Tyson Farrell. First in-human clinical performance of a new non-cavitating handheld lensectomy system in 665 consecutive cataract surgeries.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001446.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.