A retrievable nitinol vena cava filter: experimental and initial clinical results.

Nakagawa N, Cragg A H, Smith T P, Castaneda F, Barnhart W H, DeJong S C

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: The authors describe the properties of a new retrievable nitinol vena cava filter and report experimental and initial clinical results.

MATERIALS AND

METHODS

The filters, made of nitinol monofilament wire that forms a spiral cone and retrieval wire, were introduced through an indwelling 5.5-F transfemoral sheath into the infrarenal portion of the inferior vena cava in 10 sheep. In seven animals, four 4 x 30-mm radiopaque clots were injected below the filter to test its thrombus-trapping efficacy. Aspiration thrombectomy was then attempted, and the filter was removed. Follow-up venography was performed 1 week after placement in three other animals. After successful preclinical testing, the filter was implanted and retrieved in two patients.

RESULTS

All 10 filters were successfully and easily placed in sheep. All filters were thrombus-free at follow-up venography. All clots injected in the iliac veins were trapped by the filter and successfully removed by means of aspiration thrombectomy. All 10 filters were retrieved without difficulty. Temporary filter implantation and retrieval were accomplished in two patients for 5 and 7 days. In one patient, infrafilter thrombus was aspirated. Perisheath thrombosis occurred in both patients. One patient subsequently underwent permanent filter placement.

CONCLUSION

Temporary vena cava filtration is feasible. Potential advantages include easy placement, surveillance, and retrieval. One current limitation is pericatheter thrombosis, which may be eliminated by a less thrombogenic sheath.

Для цитирования: Nakagawa N, Cragg A H, Smith T P, Castaneda F, Barnhart W H, DeJong S C. A retrievable nitinol vena cava filter: experimental and initial clinical results.. Journal of vascular and interventional radiology : JVIR. 1994. DOI: 10.1016/s1051-0443(94)71539-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.