Automated Quantification of Decreased Fundus Autofluorescence in Stargardt Disease Using Starguage: Validation against Manual Grading Standards.

Ahmed Mohamed I, Yucel Hikmet, Afridi Rubbia, de Guimarães Thales A C, Sendino-Tenorio Isabel, Nguyen Nam V, Kiran Romaisa, Khan Sidrah, Khan Ufaq, Sharaf Un Nisa Syeda, Campigotto Mauro, Hariri Amir, Michaelides Michel, Scholl Hendrik P N, Mata Nathan, Nguyen Quan D, Sepah Yasir J

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the repeatability and reproducibility of Starguage, a novel automated method, compared with manual segmentation for measuring decreased autofluorescence (DAF) and definitely decreased autofluorescence (DDAF) in fundus autofluorescence (FAF) images of patients with Stargardt disease.

DESIGN

A cross-sectional reproducibility and agreement study.

PARTICIPANTS

A total of 316 eyes from 158 genetically confirmed Stargardt patients were analyzed. For intragrader repeatability, 114 FAF images were reassessed in a masked, repeated-measures design.

METHODS

Decreased autofluorescence and DDAF lesion areas were independently quantified by five certified graders using either manual delineation with Heidelberg RegionFinder or a threshold-based automated algorithm, with automated quantification and cross-method agreement analyses restricted to a prespecified central 6-mm fovea-centered region. Agreement and repeatability were assessed using intraclass correlation coefficients (ICCs), standard error of measurement (SEM), minimal detectable change (MDC), Lin's concordance correlation coefficient (CCC), Bland-Altman plots, and Passing-Bablok regression. Both raw and square-root-transformed lesion areas were evaluated.

MAIN OUTCOME MEASURES: Repeatability (intragrader ICC, SEM, and MDC), reproducibility (intergrader ICC), and agreement (CCC and bias in regression analysis) between and within manual and automated methods.

RESULTS

The automated method achieved excellent intragrader repeatability for both DAF and DDAF (ICCs ≥0.988, SEM ≤0.71 mm2, MDC ≤1.98 mm2), with minimal operator influence. Manual measurements showed variable repeatability (DAF ICCs 0.909-0.974; DDAF ICCs as low as 0.837), with square-root transformation reducing SEM and MDC. Intergrader reproducibility was highest for automated methods (ICC = 0.988-0.992), whereas manual methods ranged from 0.764-0.939 (raw) and 0.867-0.922 (transformed). Cross-method agreement was strong (CCC = 0.91-0.96), though minor proportional and constant bias was observed in raw DAF data.

CONCLUSIONS

The automated approach provides near-perfect repeatability and high agreement with manual grading, offering a scalable, objective alternative for quantifying hypo-auto-fluorescent lesions in Stargardt disease. Manual methods are generally reliable but more variable, especially for DDAF, and benefit from square-root transformation. Findings reflect a pediatric/adolescent single-trial predominantly Asian cohort.

FINANCIAL DISCLOSURES: The authors have no proprietary or commercial interest in any materials discussed in this article.

Для цитирования: Ahmed Mohamed I, Yucel Hikmet, Afridi Rubbia, de Guimarães Thales A C, Sendino-Tenorio Isabel, Nguyen Nam V, Kiran Romaisa, Khan Sidrah, Khan Ufaq, Sharaf Un Nisa Syeda, Campigotto Mauro, Hariri Amir, Michaelides Michel, Scholl Hendrik P N, Mata Nathan, Nguyen Quan D, Sepah Yasir J. Automated Quantification of Decreased Fundus Autofluorescence in Stargardt Disease Using Starguage: Validation against Manual Grading Standards.. Ophthalmology science. 2026. DOI: 10.1016/j.xops.2026.101080.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.