Triangular prismatic JJ stent does not cause more discomfort than tubular ones: a randomised controlled trial comparison.

Zhang Lin, Wu Yanyuan, Chen Yuangui, Qian Subo, Duan Liujian, Huang Yunteng, Ding Jie, Qi Jun, Cao Jianwei

Рандомизированное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare the stent-related symptoms (SRS) of three commonly used, readily accessible ureteric JJ stents after uncomplicated flexible ureteroscopic lithotripsy (FURL), in a prospective randomised controlled single-blind parallel-group study, in order to see whether structural difference might influence SRS.

PATIENTS AND

METHODS

Patients undergoing FURL were randomised into three groups: the Cook Group received conventional 6 F Cook Universa Soft JJ stents as control, the Kang Yi Bo (KYB) Group received 6 F KYB anti-reflux JJ stents, and the Urovision Group received 7 F Urovision Visiostar ESWL JJ stents. The ureteric stent symptom questionnaire (USSQ) was administered at 1 week, 4 weeks (before stent removal), and 5 weeks (one week after stent removal as baseline evaluation) after stent insertion. Both raw and baseline-adjusted USSQ domain subscores at 1 week and 4 weeks were compared.

RESULTS

A total of 146 patients were included in the analysis. The KYB Group showed significantly lower P6&7 subscore yet higher urinary symptoms score 1 week and 4 weeks after stents insertion than both Cook and Urovision, whilst the Urovision Group achieved similar scores in most domains with Cook.

CONCLUSIONS

Although the KYB anti-reflux JJ stent might prevent vesicoureteral reflux, it induces significantly stronger urinary symptoms, both at 1 week or 4 weeks after stent insertion, with or without baseline correction. Despite the unique triangular prismatic shape, the Urovision Visiostar stent does not cause heavier urinary symptoms or pain compared to the conventional cylinder shape counterparts.

Для цитирования: Zhang Lin, Wu Yanyuan, Chen Yuangui, Qian Subo, Duan Liujian, Huang Yunteng, Ding Jie, Qi Jun, Cao Jianwei. Triangular prismatic JJ stent does not cause more discomfort than tubular ones: a randomised controlled trial comparison.. World journal of urology. 2024. DOI: 10.1007/s00345-023-04758-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.