MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.crad.2005.06.009

Медицинская библиотека · Clinical radiology · 2005

Optimal exposure parameters for digital radiography of the infant skull: a pilot study.

Offiah A C, Grehan J, Hall C M, Todd-Pokropek A

СтатьяПеревод пока не подготовлен

Аннотация

AIM: To determine optimal exposure parameters when performing digital skull radiographs in infants with suspected non-accidental injury (NAI).

METHOD: Anteroposterior and lateral post-mortem skull radiographs of six consecutive infants with suspected NAI were made at six exposure levels for each projection. Entrance surface doses ranged from 75-351 microGy. Exposures were made with a Fuji 5000R computed radiography system onto a standard resolution imaging plate. In three patients exposures were repeated using a high-resolution imaging plate. Hard copy images with an edge-enhancement factor of 0.5 were produced. Six observers assessed and scored the radiographs from 1=poor to 5=excellent for visualization of five criteria. The criteria scored included outer table of skull vault, inner table of skull vault, suture margins, vascular markings and soft tissues of the scalp. Radiographs were then ranked in order of overall image quality. Film density and sensitivity values were recorded. Local research committee approval was obtained.

RESULTS

Current parameters give an average entrance surface dose of 253 microGy and 246 microGy for anteroposterior and lateral radiographs, respectively. The study demonstrated no perceived improvement in image quality above an entrance surface dose of 200 microGy (80% of current dose) or by the use of a high-resolution imaging plate.

CONCLUSION

The potential exists to reduce radiation exposure in infants. A study has commenced to determine the effects of dose reduction on diagnostic accuracy in suspected NAI.

Для цитирования: Offiah A C, Grehan J, Hall C M, Todd-Pokropek A. Optimal exposure parameters for digital radiography of the infant skull: a pilot study.. Clinical radiology. 2005. DOI: 10.1016/j.crad.2005.06.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.