MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.bjps.2024.01.019

Медицинская библиотека · Journal of plastic, reconstructive & aesthetic surgery : JPRAS · 2024

Scientific validation of clinical visual scales and Antera 3D™ consistency with derived measurements in the assessment of infantile haemangioma after laser therapy.

Morihisa Yoichiro, Rikimaru-Nishi Yukiko, Ohmaru Youkou, Ino Kou, Rikimaru Hideaki, Kiyokawa Kensuke

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Laser therapy is a treatment for infantile haemangiomas. The efficacy of laser therapy for red lesions is determined by visual evaluation; however, this assessment is inaccurate and lacks objectivity.

OBJECTIVE

To scientifically validate the consistency between pre- and post-treatment visual assessment grades for infantile haemangioma treated with pulsed dye laser (PDL) and the values calculated from images obtained with Antera 3D™.

METHODS

This study involved 81 cases of infantile haemangiomas treated with PDL alone from 2012 to 2015 and with Antera 3D™ images of the lesions. Using images obtained before treatment and 4-6 weeks after the last treatment, the lesions were rated using a visual four-step scale. Ratings were categorised as Poor/Fair/Good/Excellent by the degree of improvement in the red colour tone. The red colour ratio was calculated using the haemoglobin distribution in the lesion and surrounding skin, and the improvement difference and improvement rate were then obtained. The correlation between the improvement difference and improvement rate, and visual evaluation was statistically analysed.

RESULTS

No serious adverse effects were observed, with an average of 4.3 treatments per patient; 60.1% of the patients achieved Good/Excellent results. There were statistically significant differences in the post-treatment red colour ratio and improvement ratio in each category after visual evaluation classification. The improvement rate and the four visual grades were statistically correlated.

CONCLUSION

This study confirmed the scientific validity of visual evaluation and the evaluation criteria calculated from Antera 3D™. This method could objectively determine treatment effectiveness.

Для цитирования: Morihisa Yoichiro, Rikimaru-Nishi Yukiko, Ohmaru Youkou, Ino Kou, Rikimaru Hideaki, Kiyokawa Kensuke. Scientific validation of clinical visual scales and Antera 3D™ consistency with derived measurements in the assessment of infantile haemangioma after laser therapy.. Journal of plastic, reconstructive & aesthetic surgery : JPRAS. 2024. DOI: 10.1016/j.bjps.2024.01.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.