DOI 10.3928/15428877-20120301-01
Assessment of intrastromal corneal ring segment position with anterior segment optical coherence tomography.
Аннотация
BACKGROUND AND
OBJECTIVE
To examine the exact position of implanted Keraring (Mediphacos, Belo Horizonte, Brazil) segments by high-resolution anterior segment optical coherence tomography (AS-OCT).
PATIENTS AND
METHODS
This study included 17 eyes of 13 patients with keratoconus who underwent uneventful intracorneal ring segment implantation with the aid of femtosecond laser. Eyes were evaluated by AS-OCT at the third postoperative month. Distance from the apex of the triangular cross-section of the ring segment to the anterior corneal surface and distances from two basal corners to the posterior corneal surface were measured.
RESULTS
Distance from apex to anterior corneal surface (263.1 ± 42.9 μm) was significantly smaller than target depth calculated intraoperatively (356.9 ± 35.7 μm, P < .001). Distance between outer basal corner and posterior corneal surface was significantly greater than distance between inner basal corner and posterior corneal surface.
CONCLUSION
High-resolution AS-OCT is a rapid, convenient, and valuable technique in the follow-up of patients with implanted Keraring segments that may be helpful in predicting ring-related complications.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.