MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1364/boe.2.003037

Медицинская библиотека · Biomedical optics express · 2011

Micrometer axial resolution OCT for corneal imaging.

Yadav Rahul, Lee Kye-Sung, Rolland Jannick P, Zavislan James M, Aquavella James V, Yoon Geunyoung

СтатьяПеревод пока не подготовлен

Аннотация

An optical coherence tomography (OCT) for high axial resolution corneal imaging is presented. The system uses 375 nm bandwidth (625 to 1000 nm) from a broadband supercontinuum light source. The system was developed in free space to minimize image quality degradation due to dispersion. A custom-designed spectrometer based on a Czerny Turner configuration was implemented to achieve an imaging depth of 1 mm. Experimentally measured axial resolution was 1.1 μm in corneal tissue and had a good agreement with the theoretically calculated resolution from the envelope of the spectral interference fringes. In vivo imaging was carried out and thin corneal layers such as the tear film and the Bowman's layer were quantified in normal, keratoconus, and contact lens wearing eyes, indicating the system's suitability for several ophthalmic applications.

Для цитирования: Yadav Rahul, Lee Kye-Sung, Rolland Jannick P, Zavislan James M, Aquavella James V, Yoon Geunyoung. Micrometer axial resolution OCT for corneal imaging.. Biomedical optics express. 2011. DOI: 10.1364/boe.2.003037.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.