MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s12265-021-10199-5

Медицинская библиотека · Journal of cardiovascular translational research · 2022

In Vitro Mitral Valve Model with Unrestricted Ventricular Access: Using Vacuum to Close the Valve and Enable Static Trans-Mitral Pressure.

Stephens Sam E, Kammien Alexander J, Paris Jacob C, Applequist Alexis P, Ingels Neil B, Jensen Hanna K, Rodgers Drew E, Cole Charles R, Wenk Jonathan F, Jensen Morten O

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

Current in vitro models of the left heart establish the pressure difference required to close the mitral valve by sealing and pressurizing the ventricular side of the valve, limiting important access to the subvalvular apparatus. This paper describes and evaluates a system that establishes physiological pressure differences across the valve using vacuum on the atrial side. The subvalvular apparatus is open to atmospheric pressure and accessible by tools and sensors, establishing a novel technique for experimentation on atrioventricular valves. Porcine mitral valves were excised and closed by vacuum within the atrial chamber. Images were used to document and analyze closure of the leaflets. Papillary muscle force and regurgitant flow rate were measured to be 4.07 N at 120 mmHg and approximately 12.1 ml/s respectively, both of which are within clinically relevant ranges. The relative ease of these measurements demonstrates the usefulness of improved ventricular access at peak pressure/force closure.

Для цитирования: Stephens Sam E, Kammien Alexander J, Paris Jacob C, Applequist Alexis P, Ingels Neil B, Jensen Hanna K, Rodgers Drew E, Cole Charles R, Wenk Jonathan F, Jensen Morten O. In Vitro Mitral Valve Model with Unrestricted Ventricular Access: Using Vacuum to Close the Valve and Enable Static Trans-Mitral Pressure.. Journal of cardiovascular translational research. 2022. DOI: 10.1007/s12265-021-10199-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.