Experimental stenting of the posterior mitral leaflet to correct prolapse in mitral valve insufficiency.

Eckstein Friedrich S, Kronen Peter W, Stalder Mario, Gygax Erich, Zobrist Claudia, Levionnois Olivier L, Jenni Hansjörg, Eberle Balthasar, Carrel Thierry P

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

This study investigated the use of a new concept of mitral valve reconstruction using a novel device to stent the posterior mitral leaflet in combination with semicircular annuloplasty. Modern mitral valve repair is an accepted modality and a routine procedure for treatment of degenerative mitral valve insufficiency. One of the most common mechanisms of mitral valve insufficiency is leaflet prolapse. In the majority of cases the posterior leaflet is dysfunctional and therapeutic reconstruction of the PII flail leaflet segment involves quadrangular resection which is usually combined to mitral annulo-plasty with a ring. A new time-saving concept of mitral valve reconstruction by stenting the posterior mitral leaflet in combination with semicircular annuloplasty is presented.

METHODS

The new mitral valve reconstruction device (Shelhigh MitroFast, Shelhigh, Inc., Union, NJ, USA) was implanted in four adult sheep. It is constructed as an annuloplasty ring in combination with a posterior leaflet stent. The device has the shape of a closed posterior leaflet and forms a "buttress" against which the anterior leaflet can coapt.

RESULTS

Every implantation of a MitroFast device could be performed in less than 30 minutes. After implantation of the device, all animals could be successfully weaned from CPB. Invasively measured left atrial pressure was below 12 mm Hg in all animals. After chest closure, transoesophageal echocardiography revealed a competent mitral valve in all animals, without any inflow restriction in three animals, and suspected mild stenosis in one animal.

CONCLUSIONS

In this experimental model, implantation of the newly designed annuloplasty ring with stenting the posterior mitral leaflet avoids extensive and time-consuming reconstructive surgery on a flail posterior leaflet. Implantation of the device resulted in favorable short-term hemodynamic effects. Implantation technique of the device is simple, the potential for minimal invasive implantation of a conceptual similar device will be further investigated.

Для цитирования: Eckstein Friedrich S, Kronen Peter W, Stalder Mario, Gygax Erich, Zobrist Claudia, Levionnois Olivier L, Jenni Hansjörg, Eberle Balthasar, Carrel Thierry P. Experimental stenting of the posterior mitral leaflet to correct prolapse in mitral valve insufficiency.. The heart surgery forum. 2005. DOI: 10.1532/hsf98.20041175.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.