Dynamic Cyclic Fatigue Resistance and Bending Stiffness of Blue-Treated NiTi Rotary Instruments at Simulated Intracanal Temperature: A Comparative In Vitro Study.

Al Omari Taher, Al Jadaa Anas, Al-Zain Yazan, Abdulazziz Haytham, Atmeh Amre R, El Abed Rashid

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

To compare the dynamic cyclic fatigue resistance, bending stiffness and phase transformation characteristics of blue heat-treated instruments, VDW Rotate (VR) and Azur Blue (AzB), at a simulated intracanal temperature of 34 ± 1°C.

METHODS

Fifty-eight new instruments (n=29 per group) were allocated as follows: dynamic cyclic fatigue testing (n=12), bending stiffness evaluation (n=12), differential scanning calorimetry (DSC) (n=1), X-ray diffraction (XRD) (n=2) and energy-dispersive X-ray fluorescence (EDXRF) (n=2). Dynamic cyclic fatigue testing was conducted in an artificial canal (18 mm length, 1.4 mm diameter, 45° curvature, 5 mm radius) at 34 ± 1°C. Time to fracture was recorded and converted to the number of cycles to failure, and fractured segment lengths were measured. Bending stiffness was evaluated with a 3 mm deflection range at 34 ± 1°C. Differential scanning calorimetry, energy-dispersive X-ray fluorescence, X-ray diffraction and scanning electron microscopy were used for metallurgical and fractographic analyses.

RESULTS

Differential scanning calorimetry revealed austenitic finish temperatures (AuF) of 35.8°C for VDW Rotate and 39.4°C for AzB. Azur Blue showed significantly higher cyclic fatigue resistance (2039 ± 250 vs. 1609 ± 222 cycles), lower bending stiffness (0.85 ± 0.07 vs. 1.07 ± 0.08 N) and longer fractured fragment length (4.87± 0.44 vs. 3.27 ± 0.75 mm) than VDW Rotate (P < .05).

CONCLUSION

Azur Blue showed higher cyclic fatigue resistance, whereas VDW Rotate showed higher bending stiffness. These differences were more consistently associated with phase transformation behaviour than with external geometry alone.

Для цитирования: Al Omari Taher, Al Jadaa Anas, Al-Zain Yazan, Abdulazziz Haytham, Atmeh Amre R, El Abed Rashid. Dynamic Cyclic Fatigue Resistance and Bending Stiffness of Blue-Treated NiTi Rotary Instruments at Simulated Intracanal Temperature: A Comparative In Vitro Study.. European endodontic journal. 2026. DOI: 10.65717/eej.2026.26062.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.