Journal of cardiovascular electrophysiology · 2020

Electromagnetic interference from left ventricular assist devices in patients with subcutaneous implantable cardioverter-defibrillators.

Клинический случайСтатьяОбзорНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ishida Yuji, Payne Joshua E, Field Michael E, Gold Michael R

Аннотация

INTRODUCTION: Interactions of left ventricular assist devices (LVADs) with transvenous implantable cardioverter-defibrillator systems (ICDs) have been widely reported. However, less is known regarding the impact of electromagnetic interference (EMI) from LVADs on subcutaneous ICD function.

METHODS AND RESULTS: A comprehensive literature search was performed on PubMed, Cochrane central registry, and Google Scholar using the search terms "subcutaneous implantable cardioverter-defibrillator and left ventricular assist devices," "electromagnetic interference, LVAD, and subcutaneous ICD," "EMI and S-ICD," and "inappropriate shocks, LVAD, and ICD." Demographic and programming data were extracted from the reports and authors as needed. A total of seven cases of EMI in LVAD patients with subcutaneous ICD (S-ICD) devices were found. In addition three previously unreported cases from our center were included. All cases involved either a heartware ventricular assist device or HeartMate III LVAD with a pre-existing S-ICD. In all patients, both the primary and secondary vectors had inappropriate sensing due to EMI. Three patients were reprogramed to the alternate vector with appropriate sensing. The S-ICD was either inactivated or replaced with a transvenous device in six patients. A single patient was left sensing in the alternate vector. There were no reports of inability to interrogate S-ICD systems in patients with LVADs.

CONCLUSION: The risk of inappropriate shocks from LVADs should be considered in pre-existing patients with S-ICD, particularly when the heartware ventricular assist device or HeartMate III LVAD device is present. Reprogramming of the sensing vector can occasionally avoid this issue but often the S-ICD needs to be inactivated.