Optimizing Measurement and Interpretation of the G6PD/Hb Ratio.

Higgins Victoria, Cheng Pow Lee, Selvaratnam Rajeevan, Brinc Davor

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Glucose-6-phosphate dehydrogenase (G6PD)/hemoglobin (Hb) ratio helps detect G6PD deficiency, an X-linked disorder that can be asymptomatic or cause acute hemolytic anemia and chronic hemolysis. We investigated preanalytical, analytical, and postanalytical aspects to optimize G6PD/Hb measurement and interpretation.

METHODS

G6PD was measured with the Pointe Scientific assay and Hb with Drabkin's reagent on Alinity c® (Abbott Diagnostics). Stability of G6PD/Hb was assessed after 7 and 14 days while stored at 2-8 °C. Stability of hemolysate prepared for G6PD analysis was assessed using QC and patient samples up to 4 h at room temperature or 2-8 °C. Analytical performance specifications including precision, method comparison, linearity, LOQ, and carry-over were established for the enzymatic reaction of G6PD and spectrophotometric reading of Hb. G6PD/Hb reference interval and cut-offs were established indirectly using truncated maximum likelihood method (TML) using retrospective data (n = 4715 patient data points).

RESULTS

Samples were stable after 7 days at 2-8°C, unless grossly hemolyzed. Hemolysate prepared for G6PD measurement remained stable for up to 4 h for QC at room temperature and 2-8°C, but up to 30 min-1 h at room temperature and 1-2 h at 2-8 °C for patient samples. Precision, linearity, LOQ, and carryover were acceptable. G6PD/Hb cut-offs were <3.3, ≥3.3, 3.3-8.9, and ≥8.9 U/g Hb for deficient males/females, normal males, intermediate females, and normal females, respectively.

CONCLUSIONS

In vitro hemolysis and delayed hemolysate analysis significantly reduce G6PD/Hb stability. QC material cannot detect the impact of delayed hemolysate analysis. These findings were foundational for optimizing G6PD/Hb protocols for a new platform and establishing laboratory-specific G6PD/Hb cut-offs.

Для цитирования: Higgins Victoria, Cheng Pow Lee, Selvaratnam Rajeevan, Brinc Davor. Optimizing Measurement and Interpretation of the G6PD/Hb Ratio.. The journal of applied laboratory medicine. 2021. DOI: 10.1093/jalm/jfab008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.