MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/eye.2013.209

Медицинская библиотека · Eye (London, England) · 2013

The effect of head tilt on keratometric measurement using the IOLMaster.

Park C Y, Do J R, Kim S H, Lim C-Y, Chuck R S

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the effect of head tilt on keratometric measurement using the IOLMaster.

METHODS

Twenty-seven right eyes of 27 volunteers were examined using a manual keratometer (MK), automated keratometer (AK), and an IOLMaster. MK and AK measurements were performed in the upright head position, whereas IOLMaster measurements were performed in five different head positions (upright, 5° or 15° clockwise head tilt, and 5° or 15° counter-clockwise head tilt). The repeatability of IOLMaster was compared with MK and AK. The keratometric measurements (mean keratometric power, magnitude of astigmatism, and steep meridian) in different head positions were compared.

RESULTS

The IOLMaster showed good repeatability of keratometric measurement comparable to MK or AK. 15° and 5° clockwise head tilt resulted in 12.09 ± 9.51 (mean ± SD) (P<0.001 vs upright) and 5.51 ± 5.97 (mean ± SD) degrees (P<0.001) of clockwise rotation of steep meridian, respectively. 15° and 5° counter-clockwise head tilt resulted in 12.49 ± 7.07 (mean ± SD) (P<0.001) and 6.08 ± 5.09 (mean ± SD) degrees (P<0.001) of counter-clockwise rotation of steep meridian, respectively.

CONCLUSIONS

The patient's head tilt (5° or 15°, clockwise or counter-clockwise) significantly changed the steep meridian of astigmatism measured using IOLMaster and the steep meridian generally shifted to the direction of head tilt. The importance of maintaining the upright head posture during IOLMaster measurement is demonstrated.

Для цитирования: Park C Y, Do J R, Kim S H, Lim C-Y, Chuck R S. The effect of head tilt on keratometric measurement using the IOLMaster.. Eye (London, England). 2013. DOI: 10.1038/eye.2013.209.