Early outcomes of aortic valve replacement with Perceval PLUS sutureless valve: results of the prospective multicentric MANTRA study.

Micovic Slobodan, Nobre Angelo, Choi Jae Woong, Solinas Marco, Shehada Sharaf-Eldin, Torella Michele, Baeza Cristian, Parrino Eugene, Pollari Francesco, Troise Giovanni, Kappert Utz, Mellert Friedrich, Je Hyung Gon, Argano Vincenzo, Lam Ka Yan, Rinaldi Mauro, Gutermann Herbert, Meuris Bart, MANTRA Investigators

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The aim of this study is to report the preliminary real-word clinical and hemodynamic performance from the MANTRA study in patients undergoing aortic valve replacement with Perceval PLUS sutureless valve.

METHODS

MANTRA is an ongoing "umbrella" prospective, multi-center, international post-market study to collect real-life safety and performance data on Corcym devices (Corcym S.r.l, Saluggia, Italy). Clinical and echocardiographic outcomes were collected preoperatively, at discharge and at each follow up. KCCQ-12 and EQ-5D-5L quality of life questionnaires were collected preoperatively and at 30-days.

RESULTS

A total of 328 patients underwent aortic valve replacement with Perceval PLUS in 29 International institutions. Patients were enrolled from July 2021 to October 2023 and enrollment is still ongoing. Mean age was 71.9 ± 6.4 years, mean EuroSCORE II was 2.9 ± 3.9. Minimally invasive approach was performed in 44.2% (145/328) of patients; concomitant procedures were done in 40.8% (134/328) of cases. Thirty-day mortality was 1.8% (6/328) and no re-interventions were reported. Pacemaker implant was required in 4.0% (13/328) of the patients. The assessment of the functional status demonstrated marked and stable improvement in NYHA class in most patients at 30-day follow-up, with significant increase of KCCQ-12 summary score (from 58.8 ± 23.0 to 71.8 ± 22.1, p < 0.0001) and EQ-5D-5L VAS score (from 64.5 ± 20.4 to 72.6 ± 17.5, p < 0.0001). Mean pressure gradient decreased from 46.2 ± 17.3 mmHg to 10.1 ± 4.7 mmHg at 30-day follow-up. Low or no incidence of moderate-to-severe paravalvular or central leak was reported.

CONCLUSIONS

Preliminary results demonstrate good clinical outcomes and significant improvement of Quality of Life at 30-days, excellent early hemodynamic performance within patient implanted with Perceval PLUS.

TRIAL REGISTRATION: The MANTRA study has been registered in ClinicalTrials.gov (NCT05002543, Initial release 26 July 2021).

Для цитирования: Micovic Slobodan, Nobre Angelo, Choi Jae Woong, Solinas Marco, Shehada Sharaf-Eldin, Torella Michele, Baeza Cristian, Parrino Eugene, Pollari Francesco, Troise Giovanni, Kappert Utz, Mellert Friedrich, Je Hyung Gon, Argano Vincenzo, Lam Ka Yan, Rinaldi Mauro, Gutermann Herbert, Meuris Bart, MANTRA Investigators. Early outcomes of aortic valve replacement with Perceval PLUS sutureless valve: results of the prospective multicentric MANTRA study.. Journal of cardiothoracic surgery. 2024. DOI: 10.1186/s13019-024-02861-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.