Temporal trends with the Evolut family of self-expanding transcatheter heart valves: A single-center experience.

Merdler Ilan, Case Brian C, Bhogal Sukhdeep, Reddy Pavan K, Zhang Cheng, Ali Syed, Gallino Paige E, Jackman Caroline, Ben-Dor Itsik, Satler Lowell F, Cohen Jeffery E, Rogers Toby, Waksman Ron

СтатьяСравнительное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The Evolut self-expanding valve (SEV) systems (Medtronic), were designed to accommodate varying valve sizes and reduce paravalvular leak (PVL) while maintaining a low delivery profile. These systems have evolved between product generations, alongside valve deployment techniques changing over time.

AIMS: This study aimed to examine whether these changes impacted clinical outcomes.

METHODS

EPROMPT is a prospective, investigator-initiated, postmarketing registry of consecutive patients undergoing transfemoral transcatheter aortic valve replacement (TAVR) using the Evolut PRO/PRO+ SEV system. A total of 300 patients were divided into three consecutive cohorts of 100 patients according to implantation date (January to October 2018, November 2018 to July 2020, and August 2020 to November 2021). Procedural and clinical outcomes over these time periods were compared.

RESULTS

Valve Academic Research Consortium (VARC)-2 device implantation success improved over time (70.0% vs. 78.0% vs. 88.8%, p = 0.01), with a similar trend for VARC-3 device success (94.7% vs. 81.7% vs. 96.8%, p < 0.001). PVL (all degrees) frequency was likewise reduced over time (31.0% vs. 17.0% vs. 19.2%, p = 0.04). Furthermore, a trend was noticed toward shorter procedure times and shorter length of stay. However, postprocedural pacemaker implantation rates did not significantly differ (15.2% vs. 21.1% vs. 14.0%, p = 0.43).

CONCLUSION

During a 3-year period, we demonstrated better TAVR outcomes with newer SEV iterations, alongside changes in implantation techniques, which might result in better procedural and clinical outcomes. However, we did not see a significant change in peri-procedural pacemaker rates for SEV.

Для цитирования: Merdler Ilan, Case Brian C, Bhogal Sukhdeep, Reddy Pavan K, Zhang Cheng, Ali Syed, Gallino Paige E, Jackman Caroline, Ben-Dor Itsik, Satler Lowell F, Cohen Jeffery E, Rogers Toby, Waksman Ron. Temporal trends with the Evolut family of self-expanding transcatheter heart valves: A single-center experience.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2024. DOI: 10.1002/ccd.31088.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.