Near-real time oculodynamic MRI: a feasibility study for evaluation of diplopia in comparison with clinical testing.

Berg Isabelle, Palmowski-Wolfe Anja, Schwenzer-Zimmerer Katja, Kober Cornelia, Radue Ernst-Wilhelm, Zeilhofer Hans-Florian, Scheffler Klaus, Kunz Christoph, Buitrago-Tellez Carlos

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To demonstrate feasibility of near-real-time oculodynamic magnetic resonance imaging (od-MRI) in depicting extraocular muscles and correlate quantitatively the motion degree in comparison with clinical testing in patients with diplopia.

METHODS

In 30 od-MRIs eye movements were tracked in the horizontal and sagittal plane using a a TrueFISP sequence with high temporal resolution. Three physicians graded the visibility of extraocular muscles by a qualitative scale. In 12 cases, the maximal monocular excursions in the horizontal and vertical direction of both eyes were measured in od-MRIs and a clinical test and correlated by the Pearson test.

RESULTS

The medial and lateral rectus muscles were visible in the axial plane in 93% of the cases. The oblique, superior and inferior rectus muscles were overall only in 14% visible. Horizontal (p = 0,015) and vertical (p = 0,029) movements of the right eye and vertical movement of the left eye (p = 0,026) measured by od-MRI correlated positively to the clinical measurements.

CONCLUSIONS

Od-MRI is a feasible technique. Visualization of the horizontal/vertical rectus muscles is better than for the superior/inferior oblique muscle. Od-MRI correlates well with clinical testing and may reproduce the extent of eye bulb motility and extraocular muscle structural or functional deteriorations. Key Points • Oculodynamic MRI technique helps clinicians to assess eye bulb motility disorders • MRI evaluation of eye movement provides functional information in cases of diplopia • Oculodynamic MRI reproduces excursion of extraocular muscles with good correlation with clinical testing • Dynamic MRI sequence supplements static orbital protocol for evaluation of motility disorders.

Для цитирования: Berg Isabelle, Palmowski-Wolfe Anja, Schwenzer-Zimmerer Katja, Kober Cornelia, Radue Ernst-Wilhelm, Zeilhofer Hans-Florian, Scheffler Klaus, Kunz Christoph, Buitrago-Tellez Carlos. Near-real time oculodynamic MRI: a feasibility study for evaluation of diplopia in comparison with clinical testing.. European radiology. 2011. DOI: 10.1007/s00330-011-2232-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.