Early Stent Frame Fracture of SAPIEN 3 Valve in the Pulmonary Position.

Haddad Raymond N, Batteux Clément, Belli Emre, Hascoët Sébastien

СтатьяКлинический случайVideo-Audio MediaПеревод в очереди

Аннотация

A 15-year-old post-REV patient underwent transcatheter pulmonary valve replacement with a 23 mm SAPIEN 3 for right ventricular outflow tract (RVOT) stenosis and severe regurgitation. At 9 months, recurrent stenosis revealed multiple frame fractures. Balloon redilation up to 24 mm was limited by aortic root compression and recoil, prompting surgery. This first reported case of early SAPIEN 3 fracture in the pulmonary position highlights structural risks in complex RVOT anatomies.

Для цитирования: Haddad Raymond N, Batteux Clément, Belli Emre, Hascoët Sébastien. Early Stent Frame Fracture of SAPIEN 3 Valve in the Pulmonary Position.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2025. DOI: 10.1002/ccd.70010.