Transcatheter Aortic Valve Replacement in Pure Native Aortic Valve Regurgitation.

Yoon Sung-Han, Schmidt Tobias, Bleiziffer Sabine, Schofer Niklas, Fiorina Claudia, Munoz-Garcia Antonio J, Yzeiraj Ermela, Amat-Santos Ignacio J, Tchetche Didier, Jung Christian, Fujita Buntaro, Mangieri Antonio, Deutsch Marcus-Andre, Ubben Timm, Deuschl Florian, Kuwata Shingo, De Biase Chiara, Williams Timothy, Dhoble Abhijeet, Kim Won-Keun, Ferrari Enrico, Barbanti Marco, Vollema E Mara, Miceli Antonio, Giannini Cristina, Attizzani Guiherme F, Kong William K F, Gutierrez-Ibanes Enrique, Jimenez Diaz Victor Alfonso, Wijeysundera Harindra C, Kaneko Hidehiro, Chakravarty Tarun, Makar Moody, Sievert Horst, Hengstenberg Christian, Prendergast Bernard D, Vincent Flavien, Abdel-Wahab Mohamed, Nombela-Franco Luis, Silaschi Miriam, Tarantini Giuseppe, Butter Christian, Ensminger Stephan M, Hildick-Smith David, Petronio Anna Sonia, Yin Wei-Hsian, De Marco Federico, Testa Luca, Van Mieghem Nicolas M, Whisenant Brian K, Kuck Karl-Heinz, Colombo Antonio, Kar Saibal, Moris Cesar, Delgado Victoria, Maisano Francesco, Nietlispach Fabian, Mack Michael J, Schofer Joachim, Schaefer Ulrich, Bax Jeroen J, Frerker Christian, Latib Azeem, Makkar Raj R

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Limited data exist about safety and efficacy of transcatheter aortic valve replacement (TAVR) in patients with pure native aortic regurgitation (AR).

OBJECTIVES

This study sought to compare the outcomes of TAVR with early- and new-generation devices in symptomatic patients with pure native AR.

METHODS

From the pure native AR TAVR multicenter registry, procedural and clinical outcomes were assessed according to VARC-2 criteria and compared between early- and new-generation devices.

RESULTS

A total of 331 patients with a mean STS score of 6.7 ± 6.7 underwent TAVR. The early- and new-generation devices were used in 119 patients (36.0%) and 212 patients (64.0%), respectively. STS score tended to be lower in the new-generation device group (6.2 ± 6.7 vs. 7.6 ± 6.7; p = 0.08), but transfemoral access was more frequently used in the early-generation device group (87.4% vs. 60.8%; p < 0.001). Compared with the early-generation devices, the new-generation devices were associated with a significantly higher device success rate (81.1% vs. 61.3%; p < 0.001) due to lower rates of second valve implantation (12.7% vs. 24.4%; p = 0.007) and post-procedural AR ≥ moderate (4.2% vs. 18.8%; p < 0.001). There were no significant differences in major 30-day endpoints between the 2 groups. The cumulative rates of all-cause and cardiovascular death at 1-year follow-up were 24.1% and 15.6%, respectively. The 1-year all-cause mortality rate was significantly higher in the patients with post-procedural AR ≥ moderate compared with those with post-procedural AR ≤ mild (46.1% vs. 21.8%; log-rank p = 0.001). On multivariable analysis, post-procedural AR ≥ moderate was independently associated with 1-year all-cause mortality (hazard ratio: 2.85; 95% confidence interval: 1.52 to 5.35; p = 0.001).

CONCLUSIONS

Compared with the early-generation devices, TAVR using the new-generation devices was associated with improved procedural outcomes in treating patients with pure native AR. In patients with pure native AR, significant post-procedural AR was independently associated with increased mortality.

Для цитирования: Yoon Sung-Han, Schmidt Tobias, Bleiziffer Sabine, Schofer Niklas, Fiorina Claudia, Munoz-Garcia Antonio J, Yzeiraj Ermela, Amat-Santos Ignacio J, Tchetche Didier, Jung Christian, Fujita Buntaro, Mangieri Antonio, Deutsch Marcus-Andre, Ubben Timm, Deuschl Florian, Kuwata Shingo, De Biase Chiara, Williams Timothy, Dhoble Abhijeet, Kim Won-Keun, Ferrari Enrico, Barbanti Marco, Vollema E Mara, Miceli Antonio, Giannini Cristina, Attizzani Guiherme F, Kong William K F, Gutierrez-Ibanes Enrique, Jimenez Diaz Victor Alfonso, Wijeysundera Harindra C, Kaneko Hidehiro, Chakravarty Tarun, Makar Moody, Sievert Horst, Hengstenberg Christian, Prendergast Bernard D, Vincent Flavien, Abdel-Wahab Mohamed, Nombela-Franco Luis, Silaschi Miriam, Tarantini Giuseppe, Butter Christian, Ensminger Stephan M, Hildick-Smith David, Petronio Anna Sonia, Yin Wei-Hsian, De Marco Federico, Testa Luca, Van Mieghem Nicolas M, Whisenant Brian K, Kuck Karl-Heinz, Colombo Antonio, Kar Saibal, Moris Cesar, Delgado Victoria, Maisano Francesco, Nietlispach Fabian, Mack Michael J, Schofer Joachim, Schaefer Ulrich, Bax Jeroen J, Frerker Christian, Latib Azeem, Makkar Raj R. Transcatheter Aortic Valve Replacement in Pure Native Aortic Valve Regurgitation.. Journal of the American College of Cardiology. 2017. DOI: 10.1016/j.jacc.2017.10.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.