Qian Zhen, Wang Kan, Liu Shizhen, Zhou Xiao, Rajagopal Vivek, Meduri Christopher, Kauten James R, Chang Yung-Hang, Wu Changsheng, Zhang Chuck, Wang Ben, Vannan Mani A
Аннотация
OBJECTIVES: This study aimed to develop a procedure simulation platform for in vitro transcatheter aortic valve replacement (TAVR) using patient-specific 3-dimensional (3D) printed tissue-mimicking phantoms. We investigated the feasibility of using these 3D printed phantoms to quantitatively predict the occurrence, severity, and location of any degree of post-TAVR paravalvular leaks (PVL).
BACKGROUND: We have previously shown that metamaterial 3D printing technique can be used to create patient-specific phantoms that mimic the mechanical properties of biological tissue. This may have applications in procedural planning for cardiovascular interventions.
METHODS: This retrospective study looked at 18 patients who underwent TAVR. Patient-specific aortic root phantoms were created using the tissue-mimicking 3D printing technique using pre-TAVR computed tomography. The CoreValve (self-expanding valve) prostheses were deployed in the phantoms to simulate the TAVR procedure, from which post-TAVR aortic root strain was quantified in vitro. A novel index, the annular bulge index, was measured to assess the post-TAVR annular strain unevenness in the phantoms. We tested the comparative predictive value of the bulge index and other known predictors of post-TAVR PVL.
RESULTS: The maximum annular bulge index was significantly different among patient subgroups that had no PVL, trace-to-mild PVL, and moderate-to-severe PVL (p = 0.001). Compared with other known PVL predictors, bulge index was the only significant predictor of moderate-severe PVL (area under the curve = 95%; p < 0.0001). Also, in 12 patients with post-TAVR PVL, the annular bulge index predicted the major PVL location in 9 patients (accuracy = 75%).
CONCLUSIONS: In this proof-of-concept study, we have demonstrated the feasibility of using 3D printed tissue-mimicking phantoms to quantitatively assess the post-TAVR aortic root strain in vitro. A novel indicator of the post-TAVR annular strain unevenness, the annular bulge index, outperformed the other established variables and achieved a high level of accuracy in predicting post-TAVR PVL, in terms of its occurrence, severity, and location.
BACKGROUND: We have previously shown that metamaterial 3D printing technique can be used to create patient-specific phantoms that mimic the mechanical properties of biological tissue. This may have applications in procedural planning for cardiovascular interventions.
METHODS: This retrospective study looked at 18 patients who underwent TAVR. Patient-specific aortic root phantoms were created using the tissue-mimicking 3D printing technique using pre-TAVR computed tomography. The CoreValve (self-expanding valve) prostheses were deployed in the phantoms to simulate the TAVR procedure, from which post-TAVR aortic root strain was quantified in vitro. A novel index, the annular bulge index, was measured to assess the post-TAVR annular strain unevenness in the phantoms. We tested the comparative predictive value of the bulge index and other known predictors of post-TAVR PVL.
RESULTS: The maximum annular bulge index was significantly different among patient subgroups that had no PVL, trace-to-mild PVL, and moderate-to-severe PVL (p = 0.001). Compared with other known PVL predictors, bulge index was the only significant predictor of moderate-severe PVL (area under the curve = 95%; p < 0.0001). Also, in 12 patients with post-TAVR PVL, the annular bulge index predicted the major PVL location in 9 patients (accuracy = 75%).
CONCLUSIONS: In this proof-of-concept study, we have demonstrated the feasibility of using 3D printed tissue-mimicking phantoms to quantitatively assess the post-TAVR aortic root strain in vitro. A novel indicator of the post-TAVR annular strain unevenness, the annular bulge index, outperformed the other established variables and achieved a high level of accuracy in predicting post-TAVR PVL, in terms of its occurrence, severity, and location.
Предметные рубрики MeSH
AgedAged, 80 and overAortic ValveAortic Valve InsufficiencyAortic Valve StenosisFeasibility StudiesFemaleHeart Valve ProsthesisHemodynamicsHumansMaleModels, CardiovascularPatient-Specific ModelingPhantoms, ImagingPredictive Value of TestsPrinting, Three-DimensionalProof of Concept StudyProsthesis DesignRadiographic Image Interpretation, Computer-AssistedRetrospective StudiesRisk AssessmentRisk FactorsStress, MechanicalTomography, X-Ray ComputedTranscatheter Aortic Valve ReplacementTreatment Outcome