DOI 10.1093/ejcts/ezx066

Медицинская библиотека · European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2017

Safety, effectiveness and haemodynamic performance of a new stented aortic valve bioprosthesis.

Klautz Robert J M, Kappetein A Pieter, Lange Rüdiger, Dagenais Francois, Labrousse Louis, Bapat Vinayak, Moront Michael, Misfeld Martin, Zeng Cathy, Sabik Iii Joseph F, PERIGON Investigators

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

We assessed the safety, effectiveness and haemodynamic performance of a new bovine stented aortic valve bioprosthesis (Avalus™).

METHODS

The PERIGON Pivotal Trial is a prospective, non-randomized, multicentre study. Subjects had symptomatic moderate or severe aortic stenosis or chronic, severe aortic regurgitation. Death, valve-related adverse events (AEs), functional recovery and haemodynamic performance were assessed at discharge, 3-6 months and 1 year. The primary analysis compared 'late' (>30 days post-implant) linearized rates of valve-related thromboembolism, thrombosis, all and major haemorrhage, all and major paravalvular leak (PVL) and endocarditis after implantation with objective performance criteria (OPC) for AEs, in accordance with EN ISO 5840:2009. We hypothesized that the upper 95% confidence bounds of the true linearized AE rates would be ≥ 2 × OPC; rejection of the null hypothesis would demonstrate that these rates were below acceptable rates. The analysis was required to include at least 150 patients followed to 1 year and 400 valve-years. Kaplan-Meier survival analysis was also performed.

RESULTS

Total number of valve-years was 459.5 (n = 686). Linearized rates were <2 × OPC for death and valve-related thromboembolism, valve thrombosis, all and major PVL, and endocarditis, but ≥2 × OPC for all and major haemorrhage. Survival at 1 year (n = 270) was 96.4%. Patients showed good functional recovery, and haemodynamic performance was within expected range.

CONCLUSIONS

This analysis demonstrated a good safety profile and clinical effectiveness of the Avalus valve except for bleeding rates. The linearized rates of all and major haemorrhage may be related to long-term anticoagulation for non-valvular indications and the length of follow-up of this cohort.

TRIAL REGISTRATION: NCT02088554 (www.clinicaltrials.gov).

Для цитирования: Klautz Robert J M, Kappetein A Pieter, Lange Rüdiger, Dagenais Francois, Labrousse Louis, Bapat Vinayak, Moront Michael, Misfeld Martin, Zeng Cathy, Sabik Iii Joseph F, PERIGON Investigators. Safety, effectiveness and haemodynamic performance of a new stented aortic valve bioprosthesis.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2017. DOI: 10.1093/ejcts/ezx066.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.