OBJECTIVES

The purpose of this study was to determine the mechanisms of sudden death (SD) in patients with ventricular tachyarrhythmias (ventricular tachycardia/ventricular fibrillation [VT/VF]) treated with an implantable cardioverter defibrillator (ICD).

BACKGROUND

Despite ICD therapy, some patients with VT/VF still die suddenly. Optimal ICD use requires determination of the mechanisms of these residual SDs.

METHODS

We reviewed 320 patient deaths during trials of Medtronic transvenous ICD systems (Medtronic Inc., Minneapolis, Minnesota). Sudden deaths were further categorized according to mechanism. Post-shock electromechanical dissociation (EMD) describes a scenario where VT/VF was appropriately detected and treated by an ICD shock that restored a physiologic rhythm, but death still occurred immediately by EMD.

RESULTS

A mode of death could be ascribed for 317 patients-90 (28%) were sudden, 156 (49%) were non-sudden cardiac, and 71 (22%) were noncardiac. A mechanism of SD was proposed for 68 patients-20 (29%) had post-shock EMD, 17 (25%) had VT/VF uncorrected by shocks, 11 (16%) had primary electromechanical dissociation, 9 (13%) had incessant VT/VF, 5 (7%) had VT/VF after their ICD was deactivated or removed, and 6 (9%) had single instances of various other terminal events. Only New York Heart Association functional class independently predicted SD by post-shock EMD.

CONCLUSIONS

The most common mechanism of SD in patients with an ICD is VT/VF treated with an appropriate shock followed by EMD. As this mechanism accounted for 29% of the SDs to which a cause could be ascribed, this mechanism of SD warrants further investigation.

Для цитирования: Mitchell L Brent, Pineda Edgar A, Titus Jack L, Bartosch Paulette M, Benditt David G. Sudden death in patients with implantable cardioverter defibrillators: the importance of post-shock electromechanical dissociation.. Journal of the American College of Cardiology. 2002. DOI: 10.1016/s0735-1097(02)01784-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.