Meier David, Akodad Mariama, Landes Uri, Barlow Aaron M, Chatfield Andrew G, Lai Althea, Tzimas Georgios, Tang Gilbert H L, Puehler Thomas, Lutter Georg, Leipsic Jonathon A, Søndergaard Lars, Wood David A, Webb John G, Sellers Stephanie L, Sathananthan Janarthanan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The implications and potential challenges of coronary access after redo transcatheter aortic valve replacement (TAVR) are unknown.

OBJECTIVES

The authors sought to evaluate the impact of different transcatheter heart valve (THV) designs, neoskirt height, implant technique, and cell misalignment on coronary access after redo TAVR.

METHODS

Different THV designs (Sapien 3 [Edwards Lifesciences LLC], Evolut Pro [Medtronic], ACURATE neo [Boston Scientific Corporation], and Portico [Abbott Structural Heart]) and sizes were implanted inside Sapien XT (Edwards Lifesciences LLC) and Evolut R (Medtronic) THVs, which were modeled as the "Ошибка" THVs, at different implant depths. Valve combinations underwent micro-computed tomography to determine the neoskirt height and dimensions of the lowest accessible cell for potential coronary access. This was compared with dimensions of 6-F/7-F/8-F coronary guiding catheters.

RESULTS

Redo TAVR combinations resulted in a wide range of neoskirt heights (15.4-31.6 mm) and a variable diameter of the lowest accessible cell (1.9-21.8 mm). An ACURATE neo implanted in a Sapien XT resulted in the largest accessible cells, whereas a Portico implanted in a Sapien XT resulted in the lowest neoskirt heights. The smallest accessible cell was observed in the Evolut Pro-in-Evolut R configuration with higher neoskirt heights. Redo TAVR in a tall frame valve with supra-annular leaflets caused a taller neoskirt height. In Evolut-in-Evolut combinations, misalignment of the cells of the 2 THVs reduced the cell area by 30% to 50% compared with an aligned configuration.

CONCLUSIONS

This study demonstrates that different redo TAVR combinations are not equivalent in terms of future coronary access. Redo TAVR using a tall frame valve in a failed tall frame valve and misaligned cells may lead to potentially challenging coronary access.

Для цитирования: Meier David, Akodad Mariama, Landes Uri, Barlow Aaron M, Chatfield Andrew G, Lai Althea, Tzimas Georgios, Tang Gilbert H L, Puehler Thomas, Lutter Georg, Leipsic Jonathon A, Søndergaard Lars, Wood David A, Webb John G, Sellers Stephanie L, Sathananthan Janarthanan. Coronary Access Following Redo TAVR: Impact of THV Design, Implant Technique, and Cell Misalignment.. JACC. Cardiovascular interventions. 2022. DOI: 10.1016/j.jcin.2022.05.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.