Combining Transcatheter Edge-to-Edge Repair and Commissural Plug Occlusion for Complex Degenerative Mitral Regurgitation.

Li Yan-Jie, Ma Lan, Hao Zi-Yong, Pan Xin

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Degenerative mitral regurgitation involving the commissures remains a challenging anatomical subset for transcatheter repair.

CASE SUMMARY: An 86-year-old male with high surgical risk presented with symptomatic, severe degenerative mitral regurgitation due to A3 prolapse extending to the posteromedial commissure and a measured mitral valve area of 3.8 cm2. The heart team was elected for transcatheter edge-to-edge repair. Despite successful implantation of a single clip in the A3-P3 region, moderate residual commissural regurgitation persisted. The small native valve area and insufficient leaflet length after the first clip precluded safe placement of a second device. As a bailout strategy, a vascular plug was deployed to specifically occlude the residual commissural leak.

DISCUSSION: This simultaneous hybrid approach successfully reduced mitral regurgitation to mild, demonstrating a novel solution for managing complex anatomical challenges where conventional transcatheter edge-to-edge repair is insufficient.

TAKE-HOME MESSAGES: Commissural involvement combined with a small mitral valve area and limited leaflet length may render conventional multiclip transcatheter edge-to-edge repair unfeasible, increasing the risk of residual regurgitation. This hybrid transcatheter strategy, combining edge-to-edge repair with targeted plug occlusion, can expand the treatable spectrum of complex degenerative mitral regurgitation in high-surgical-risk patients.

Для цитирования: Li Yan-Jie, Ma Lan, Hao Zi-Yong, Pan Xin. Combining Transcatheter Edge-to-Edge Repair and Commissural Plug Occlusion for Complex Degenerative Mitral Regurgitation.. JACC. Case reports. 2026. DOI: 10.1016/j.jaccas.2026.108350.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.