Inappropriate management of self-terminating ventricular arrhythmias by implantable cardioverter defibrillators despite a specific reconfirmation algorithm: a report of two cases.

Hurst T M, Krieglstein H, Tillmanns H, Waldecker B

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Algorithms that attempt to reconfirm the presence of an arrhythmia prior to definite treatment have been implemented in ICDs to prevent inappropriate shock therapy due to self-terminating ventricular arrhythmias. Nevertheless, in two patients, clinically inappropriate shocks were delivered after spontaneous conversion of the arrhythmia despite the use of a specific reconfirmation algorithm. Reconfirmation criteria were met due to a premature ventricular complex causing a short cycle in the first patient and a long postextrasystolic pause in the second patient. To avoid inappropriate shock therapy due to self-terminating ventricular arrhythmias, further improvement of detection algorithms is required.

Для цитирования: Hurst T M, Krieglstein H, Tillmanns H, Waldecker B. Inappropriate management of self-terminating ventricular arrhythmias by implantable cardioverter defibrillators despite a specific reconfirmation algorithm: a report of two cases.. Pacing and clinical electrophysiology : PACE. 1997. DOI: 10.1111/j.1540-8159.1997.tb06787.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.