A newly designed fully covered metal stent for lumen apposition in EUS-guided drainage and access: a feasibility study (with videos).

Moon Jong Ho, Choi Hyun Jong, Kim Dong Choon, Lee Yun Nah, Kim Hee Kyung, Jeong Song Ah, Lee Tae Hoon, Cha Sang-Woo, Cho Young Deok, Park Sang Heum, Jeong Seok, Lee Don Haeng, Isayama Hiroyuki, Itoi Takao

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАVideo-Audio MediaПеревод пока не подготовлен

Аннотация

BACKGROUND

A lumen-apposing stent can be used effectively under endosonographic guidance.

OBJECTIVE

To evaluate a newly designed, fully covered self-expandable metal stent with folding anchoring flanges for lumen apposition assembled on a conventional delivery system.

DESIGN

Retrospective case series and animal study.

SETTING

Tertiary care academic medical centers.

SUBJECTS: Six pigs for animal study and 7 patients, 3 of whom underwent endoscopic drainage for acute cholecystitis (AC) and 4 for pancreatic fluid collection (PFC).

INTERVENTION: Stent deployment under EUS guidance after puncturing, passage of an endoscope through the stent into the gallbladder (GB), or PFC with conventional endoscopic procedures.

MAIN OUTCOME MEASUREMENTS: Technical and clinical success, adverse events, and removability.

RESULTS

In the animal study, the stent was successfully inserted and deployed in the GB via a transgastric approach under EUS guidance without adverse events in all 6 pigs. Contrast injection demonstrated the absence of leakage. Cholecystoscopy with enhanced endoscopy was performed successfully in all animals after stent placement. All stents were intact and were removed successfully at 4 weeks. GB firmly adhered to the stomach with an intact cholecystogastric tract on necropsy and histopathology. The stents were successfully deployed without adverse effects in 7 patients. AC or PFC was resolved after stent placement in all patients. Endoscopic procedures were possible through the stent. Stent migration was not observed. The stent was successfully removed from the 4 patients with PFC after complete resolution.

LIMITATIONS: Small sample size, retrospective study.

CONCLUSIONS

Transenteric drainage and endoscopic intervention by using a novel fully covered self-expandable metal stent for lumen apposition under EUS guidance is feasible for the management of AC and PFC. Further study is warranted.

Для цитирования: Moon Jong Ho, Choi Hyun Jong, Kim Dong Choon, Lee Yun Nah, Kim Hee Kyung, Jeong Song Ah, Lee Tae Hoon, Cha Sang-Woo, Cho Young Deok, Park Sang Heum, Jeong Seok, Lee Don Haeng, Isayama Hiroyuki, Itoi Takao. A newly designed fully covered metal stent for lumen apposition in EUS-guided drainage and access: a feasibility study (with videos).. Gastrointestinal endoscopy. 2014. DOI: 10.1016/j.gie.2014.02.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.