Assessment of aspiration risk from dynamic modulation of endotracheal tube cuff pressure.
Аннотация
OBJECTIVES/HYPOTHESIS: To assess the risk of aspiration using a novel valve circuit that dynamically modulates endotracheal tube cuff pressure during the ventilatory cycle using bench and live animal models.
STUDY
DESIGN
Animal model.
METHODS
The bench model consisted of a cuffed endotracheal tube inserted into an artificial trachea. Leakage of liquid around the cuff was measured after 4 hours of constant or dynamic modulation of cuff pressure at variable peak end expiratory pressures. In the porcine model, eight animals were ventilated with the modulating valve circuit and compared to eight controls ventilated with a constant cuff pressure (25 cm of water). Aspiration was monitored quantitatively using a pH probe (measured as voltage) and visually using fluoroscopy.
RESULTS
There was no difference in the amount of fluid leakage around the endotracheal tube cuff in the constant or dynamically modulated pressure-cuff groups in the bench or animal models.
CONCLUSION
Dynamically modulating endotracheal tube-cuff pressures to minimize tracheal mucosal damage does not increase the tendency to leak around the cuff compared to endotracheal tube cuffs inflated to a constant pressure.
LEVEL OF EVIDENCE: N/A.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.