DOI 10.1016/j.plabm.2021.e00235

Медицинская библиотека · Practical laboratory medicine · 2021

Comparison of two immunoassays for the measurement of serum HE4 for ovarian cancer.

Barr Chloe E, Funston Garth, Mounce Luke T A, Pemberton Phillip W, Howe Jonathon D, Crosbie Emma J

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The use of Human Epididymis Protein 4 (HE4) as a biomarker for ovarian cancer is gaining traction, providing the impetus for development of a high throughput automated HE4 assay that is comparable to the conventional manual enzyme immunometric-assay (EIA). The aim of this study was to compare two immunoassay methods for the measurement of serum HE4.

MATERIALS AND

METHODS

1348 serum samples were analysed for serum HE4 using both the EIA and the automated chemiluminescent immunoassay (CLEIA) methods. HE4 values were compared using a Passing-Bablok regression and agreement assessed using Lin's concordance correlation coefficient (CCC). The absolute and percentage bias of the CLEIA compared to EIA was determined.

RESULTS

There was moderate agreement between the two methods (CCC 0.929, 95%CI 0.923-0.936). Passing-Bablok regression demonstrated an overestimation of the CLEIA [constant 4.44 (95%CI 2.96-5.68), slope 1.04 (95%CI 1.02-1.07)]. The CLEIA method had a mean percentage bias of 16.25% compared to the EIA method.

CONCLUSION

The CLEIA significantly overestimated serum HE4 values compared to the EIA, which could impact clinical interpretation and patient management. Further studies are required to develop an appropriate cut-off depending on the population being investigated and the analytic method being used.

Для цитирования: Barr Chloe E, Funston Garth, Mounce Luke T A, Pemberton Phillip W, Howe Jonathon D, Crosbie Emma J. Comparison of two immunoassays for the measurement of serum HE4 for ovarian cancer.. Practical laboratory medicine. 2021. DOI: 10.1016/j.plabm.2021.e00235.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.