DOI 10.1161/circep.114.001965

Медицинская библиотека · Circulation. Arrhythmia and electrophysiology · 2016

Novel ICD Programming and Inappropriate ICD Therapy in CRT-D Versus ICD Patients: A MADIT-RIT Sub-Study.

Kutyifa Valentina, Daubert James P, Schuger Claudio, Goldenberg Ilan, Klein Helmut, Aktas Mehmet K, McNitt Scott, Stockburger Martin, Merkely Bela, Zareba Wojciech, Moss Arthur J

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The Multicenter Automatic Defibrillator Implantation Trial-Reduce Inappropriate therapy (MADIT-RIT) trial showed a significant reduction in inappropriate implantable cardioverter defibrillator (ICD) therapy in patients programmed to high-rate cut-off (Arm B) or delayed ventricular tachycardia therapy (Arm C), compared with conventional programming (Arm A). There is limited data on the effect of cardiac resynchronization therapy with a cardioverter defibrillator (CRT-D) on the effect of ICD programming. We aimed to elucidate the effect of CRT-D on ICD programming to reduce inappropriate ICD therapy in patients implanted with CRT-D or an ICD, enrolled in MADIT-RIT.

METHODS AND

RESULTS

The primary end point of this study was the first inappropriate ICD therapy. Secondary end points were inappropriate anti-tachycardia pacing and inappropriate ICD shock. The study enrolled 742 (49%) patients with an ICD and 757 (51%) patients with a CRT-D. Patients implanted with a CRT-D had 62% lower risk of inappropriate ICD therapy than those with an ICD only (hazard ratio [HR] =0.38, 95% confidence interval: 0.25-0.57; P<0.001). High-rate cut-off or delayed ventricular tachycardia therapy programming significantly reduced the risk of inappropriate ICD therapy compared with conventional ICD programming in ICD (HR=0.14 [B versus A]; HR=0.21 [C versus A]) and CRT-D patients (HR=0.15 [B versus A]; HR=0.23 [C versus A]; P<0.001 for all). There was a significant reduction in inappropriate anti-tachycardia pacings in both group and a significant reduction in inappropriate ICD shock in CRT-D patients.

CONCLUSIONS

Patients implanted with a CRT-D have lower risk of inappropriate ICD therapy than those with an ICD. Innovative ICD programming significantly reduces the risk of inappropriate ICD therapy in both ICD and CRT-D patients.

CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov; Unique identifier: NCT00947310.

Для цитирования: Kutyifa Valentina, Daubert James P, Schuger Claudio, Goldenberg Ilan, Klein Helmut, Aktas Mehmet K, McNitt Scott, Stockburger Martin, Merkely Bela, Zareba Wojciech, Moss Arthur J. Novel ICD Programming and Inappropriate ICD Therapy in CRT-D Versus ICD Patients: A MADIT-RIT Sub-Study.. Circulation. Arrhythmia and electrophysiology. 2016. DOI: 10.1161/circep.114.001965.

Предметные рубрики MeSH