Characterization and predictors of first and subsequent inappropriate ICD therapy by heart rate ranges: Result of the MADIT-RIT efficacy analysis.

Kutyifa Valentina, Daubert James P, Olshansky Brian, Huang David T, Zhang Claire, Ruwald Anne-Christine H, McNitt Scott, Zareba Wojciech, Moss Arthur J, Schuger Claudio

СтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Data on inappropriate implantable cardioverter-defibrillator (ICD) therapy and effects of programming by heart rate are lacking.

OBJECTIVE

We aimed to characterize inappropriate ICD therapy and assess the effects of novel programming by heart rate.

METHODS

Incidence and causes of inappropriate therapy by heart rate range (below or above 200 bpm) were assessed. Predictors of inappropriate therapy and effects of programming by heart rate were evaluated with multivariate Cox regression models. Crossovers were excluded.

RESULTS

Inappropriate therapy occurred in 9.2% of the total patient population, with 19% of patients randomized to study arm A, 3.6% in arm B, and 4.7% in arm C. Inappropriate therapies <200 bpm were attributable to supraventricular tachycardia (SVT)/sinus tachycardia (78%) or atrial fibrillation/flutter (20%). Inappropriate therapy ≥200 bpm occurred because of SVT (47%), atrial fibrillation/flutter (41%), or electromagnetic interference (13%). Conventional ICD programming was associated with more inappropriate therapy <200 bpm than high-rate or delayed therapy, as were younger age, history of atrial arrhythmia, advanced New York Heart Association functional class, ICD versus cardiac resynchronization therapy with defibrillator, and absence of diabetes. High-rate and long-delay therapy significantly reduced the risk of inappropriate therapy in the <200 bpm range. Long delay was associated with further reduction of fast (≥200 bpm) inappropriate therapy (P = .032) and a reduction in subsequent inappropriate episodes (P = .006).

CONCLUSION

In MADIT-RIT, inappropriate ICD therapy is most frequent at rates below 200 bpm and can be predicted, and effectively prevented, with high-rate cutoff programming. Long-delay therapy effectively reduces fast inappropriate therapy ≥200 bpm and subsequent events. [

CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov/ct2/show/NCT00947310].

Для цитирования: Kutyifa Valentina, Daubert James P, Olshansky Brian, Huang David T, Zhang Claire, Ruwald Anne-Christine H, McNitt Scott, Zareba Wojciech, Moss Arthur J, Schuger Claudio. Characterization and predictors of first and subsequent inappropriate ICD therapy by heart rate ranges: Result of the MADIT-RIT efficacy analysis.. Heart rhythm. 2015. DOI: 10.1016/j.hrthm.2015.05.021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.