Evaluation of lower oesophageal sphincter pressure using endoscopic manometric sleeve assembly.

Hananoki M, Haruma K, Tsuga K, Hata J, Sumii K, Kajiyama G

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The purpose of this study was to establish a method to determine lower oesophageal sphincter (LOS) pressure using an endoscopic manometric sleeve assembly.

METHODS

We used a 250-cm, three-lumen, 2.2-mm external diameter silicon catheter with a 3-cm sleeve sensor that was passed easily through the biopsy channel (2.8 mm diameter) of the gastroscope. Each lumen was perfused with distilled water using a low-compliance, pneumohydraulic capillary infusion system. Forty-seven healthy subjects and 35 patients with oesophageal disorders underwent study during routine diagnostic endoscopy. In 27 of the subjects, standard transnasal manometry with a three-lumen, 4.5-mm diameter polyvinyl catheter with a 5-cm sleeve sensor was also performed.

RESULTS

Lower oesophageal sphincter pressure (LOS) was evaluated in all subjects (median pressure 16 mmHg, range 0-55 mmHg) for 6 min during routine endoscopic examination. The LOS pressure readings between standard and endoscopic manometry correlated well. The LOS pressure was significantly lower in healthy subjects with hiatal hernia than in those without hernia and was also significantly lower in patients with reflux oesophagitis than in healthy subjects.

CONCLUSIONS

Endoscopic manometric sleeve assembly was used during routine endoscopic examination and was useful for easily determining LOS pressure. It could be used for evaluation of oesophageal motility disorders, thus avoiding the use of a more invasive and time-consuming method.

Для цитирования: Hananoki M, Haruma K, Tsuga K, Hata J, Sumii K, Kajiyama G. Evaluation of lower oesophageal sphincter pressure using endoscopic manometric sleeve assembly.. Journal of gastroenterology and hepatology. 2000. DOI: 10.1046/j.1440-1746.2000.02058.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.