Low efficacy of cardioversion of persistent atrial fibrillation with the implantable cardioverter-defibrillator.

Limantoro I, Vernooy K, Weijs B, Pisters R, Debie L, Crijns H J, Blaauw Y

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Atrial fibrillation (AF) and heart failure are conditions that often coexist. Consequently, many patients with an implantable cardioverter-defibrillator (ICD) present with AF. We evaluated the effectiveness of internal cardioversion of AF in patients with an ICD.

METHODS

Retrospectively, we included 27 consecutive ICD patients with persistent AF who underwent internal cardioversion using the ICD. When ICD cardioversion failed, external cardioversion was performed.

RESULTS

Patients were predominantly male (89 %) with a mean (SD) age of 65 ± 9 years and left ventricular ejection fraction of 36 ± 17 %. Only nine (33 %) patients had successful internal cardioversion after one, two or three shocks. The remaining 18 patients underwent external cardioversion after they failed internal cardioversion, which resulted in sinus rhythm in all. A smaller left atrial volume (99 ± 36 ml vs. 146 ± 44 ml; p = 0.019), a longer right atrial cycle length (227 (186-255) vs. 169 (152-183) ms, p = 0.030), a shorter total AF history (2 (0-17) months vs. 40 (5-75) months, p = 0.025) and dual-coil ICD shock (75 % vs. 26 %, p = 0.093) were associated with successful ICD cardioversion.

CONCLUSION

Internal cardioversion of AF in ICD patients has a low success rate but may be attempted in those with small atria, a long right atrial fibrillatory cycle length and a short total AF history, especially when a dual-coil ICD is present. Otherwise, it seems reasonable to prefer external over internal cardioversion when it comes to termination of persistent AF.

Для цитирования: Limantoro I, Vernooy K, Weijs B, Pisters R, Debie L, Crijns H J, Blaauw Y. Low efficacy of cardioversion of persistent atrial fibrillation with the implantable cardioverter-defibrillator.. Netherlands heart journal : monthly journal of the Netherlands Society of Cardiology and the Netherlands Heart Foundation. 2013. DOI: 10.1007/s12471-013-0474-z.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.