MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jmsacl.2023.11.004

Медицинская библиотека · Journal of mass spectrometry and advances in the clinical lab · 2023

Use of volumetric absorptive microsampling and parallel reaction monitoring mass spectrometry for tacrolimus blood trough measurements at home in pediatric heart transplant patients.

Zhao Junfang, Setchell Kenneth D R, Zhao Xueheng, Galandi Stephanie, Garr BreAnn N, Gao Zhiqian, Chin Clifford, Stark Shelly, Steele Paul E, Ryan Thomas D

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Measurement of trough levels for calcineurin inhibitors by venipuncture sampling is a mainstay of patient management in solid organ transplant recipients but challenging in pediatric patients. Volumetric Absorptive Microsampling (VAMS) is a patient-friendly, minimally invasive sampling technique to accurately collect blood. An assay for measurement of tacrolimus in blood using VAMS, coupled with parallel reaction monitoring (PRM) mass spectrometry, was validated in pediatric heart transplant patients.

METHODS

Tacrolimus was measured by a newly developed high-resolution PRM assay and compared with low-resolution tandem mass spectrometry (MRM). Dried blood samples were collected from pediatric heart transplant patients (n = 35) using VAMS devices and a satisfaction survey was completed by patients/guardians. Tacrolimus concentrations were compared across whole liquid blood, dried blood spots, and capillary blood, and shipping stability determined.

RESULTS

The PRM assay was linear over a range 1-50 ng/mL, similar to MRM but had greater specificity due to reduced background noise. No significant differences in tacrolimus concentrations were observed between VAMS and venous blood. Tacrolimus dried on VAM tips was stable for 14 days and concentrations were unaffected by postal shipping. The variability in two simultaneously collected at-home patient samples was minimal - average concentration difference was 0.12 ± 0.94 ng/mL (p = 0.6) between paired samples.

CONCLUSION

A high resolution PRM mass spectrometry assay was developed for home-based dried blood collections for therapeutic monitoring of tacrolimus. The advantage of PRM was enhanced specificity and the VAMS devices provided a simple and convenient approach to blood sampling at home in pediatric heart transplant patients.

Для цитирования: Zhao Junfang, Setchell Kenneth D R, Zhao Xueheng, Galandi Stephanie, Garr BreAnn N, Gao Zhiqian, Chin Clifford, Stark Shelly, Steele Paul E, Ryan Thomas D. Use of volumetric absorptive microsampling and parallel reaction monitoring mass spectrometry for tacrolimus blood trough measurements at home in pediatric heart transplant patients.. Journal of mass spectrometry and advances in the clinical lab. 2023. DOI: 10.1016/j.jmsacl.2023.11.004.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.