Defining Optimal Settings for Lung Cryobiopsy in End-Stage Pulmonary Disease. A Human, Ex Vivo, Diseased Lung Clinical Trial.

Menezes Vanessa, Pollock Clare, Ferraro Pasquale, Nasir Basil, Leduc Charles, Morisset Julie, Liberman Moishe

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

To evaluate optimal settings of probe size, freezing time, and distance to the pleura that influence the size and quality of biopsy specimens during transbronchial lung cryobiopsies in ESPD.

METHODS

We prospectively recruited 17 patients undergoing lung transplantation. We created a nonperfused ex vivo bronchoscopy setting to perform multiple cryobiopsies with different probe sizes (1.7, 1.9, and 2.4 mm), freezing times (3, 5, 7, 10, 20, 30 seconds), and probe distance from pleura (5, 10, and 20 mm). Alveolated pulmonary parenchyma area≥50% in histology was considered a good quality biopsy, with a minimum procedural artifact. We used logistic regression to identify independent parameters as risk factors for histologic adequacy.

RESULTS

A total of 545 cryobiopsies were obtained from 34 explanted lungs after pneumonectomy for lung transplantation. The mean maximum diameter of the specimen achieved with the 1.7 probe was larger (13.5 mm) than those obtained with 1.9 and 2.4 mm probes (11.3 and 10.7 mm, P= 0.07). More pleural macroscopic damage and pleural tissue in histology occurred with the 2.4 mm probe ( P <0.001). There was no difference in the quality of specimens between the different freezing times and the distance from the pleura.

CONCLUSIONS

Freezing time and distance from the pleura did not affect the histologic quality for diagnosing ESPD in severely damaged lungs. Smaller cryoprobe size did not negatively affect sample adequacy.

Для цитирования: Menezes Vanessa, Pollock Clare, Ferraro Pasquale, Nasir Basil, Leduc Charles, Morisset Julie, Liberman Moishe. Defining Optimal Settings for Lung Cryobiopsy in End-Stage Pulmonary Disease. A Human, Ex Vivo, Diseased Lung Clinical Trial.. Journal of bronchology & interventional pulmonology. 2024. DOI: 10.1097/lbr.0000000000000948.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.