DOI 10.2174/1874364101105010014

Медицинская библиотека · The open ophthalmology journal · 2011

Corneal cross-linking and safety issues.

Spoerl Eberhard, Hoyer Anne, Pillunat Lutz E, Raiskup Frederik

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compile the safety aspects of the corneal collagen cross-linking (CXL) by means of the riboflavin/UVA (370 nm) approach.

MATERIALS AND METHODOLOGY: Analysis of the current treatment protocol with respect to safety during CXL.

RESULTS

The currently used UVA dose density of 5.4 J/cm(2) and the corresponding irradiance of 3 mW/cm(2) are below the known damage thresholds of UVA for the corneal endothelium, lens, and retina. Regarding the photochemical damages due to the free radicals the damage threshold for endothelial cells is 0.35 mW/cm(2). In a 400μm thick corneal stroma saturated with riboflavin, the irradiance at the endothelial level is about 0.18 mW/cm(2), which is a factor of 2 smaller than the damage threshold.

CONCLUSION

As long as the corneal stroma treated has a minimal thickness of 400 microns (as recommended), neither corneal endothelium nor deeper structures such as lens and retina will suffer any damages. The light source should provide a homogenous irradiance avoiding hot spots.

Для цитирования: Spoerl Eberhard, Hoyer Anne, Pillunat Lutz E, Raiskup Frederik. Corneal cross-linking and safety issues.. The open ophthalmology journal. 2011. DOI: 10.2174/1874364101105010014.