MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2298/sarh1006279m

Медицинская библиотека · Srpski arhiv za celokupno lekarstvo · 2010

Correlation between central corneal thickness and intraocular pressure in various age groups.

Marjanović Ivan, Kontić Djordje, Hentova-Senćanić Paraskeva, Marković Vujica, Bozić Marija

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The Pascal Dynamic Contour Tonometer (DCT) is an ophthalmic investigational device which directly calculates the dynamic pulsatile fluctuations in intraocular pressure (IOP) using a piezoelectric pressure sensor embedded in the devices tip.

OBJECTIVE

The aim of this study was to compare IOP results obtained with Goldmann applanation tonometer (GAT), and their correlation with central corneal thickness (CCT) in patients of various age groups.

METHODS

We studied 37 patients (17 male and 20 female), divided into three age groups: younger than 40 years; 40-60 years old, and older than 60 years. In the first and second group there were patients rarely and in the third mostly diagnosed and medicamentously compensated glaucoma. In topical anaesthesia, first we measured CCT, with ultrasound pachimeter, then IOP with DCT and GAT.

RESULTS

Statistically significant difference between measurements of IOP with the DCT and GAT appeared in all groups: I mean diff. -1.71 +/- 1.27 mm; p < 0.0001; II mean diff. -1.19 +/- 1.06 mm; p < 0.0001; III mean diff. -1.69 +/- 1.67 mm; p < 0.0001. CCT was in indirect correlation with the values of IOP measured both with the DCT and GAT in the first and third, while it was in the direct correlation with these values in the second group.

CONCLUSION

CCT had no influence on IOP measurements both with the DCT and GAT in none of the groups. The DCT cannot replace GAT, but it is a reliable device for the measurement of IOP particularly in corneal deformations (keratoconus, after corneal refractive surgery, corneal scars, etc.).

Для цитирования: Marjanović Ivan, Kontić Djordje, Hentova-Senćanić Paraskeva, Marković Vujica, Bozić Marija. Correlation between central corneal thickness and intraocular pressure in various age groups.. Srpski arhiv za celokupno lekarstvo. 2010. DOI: 10.2298/sarh1006279m.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.