Biomechanical and morphological corneal response to placement of intrastromal corneal ring segments for keratoconus.

Dauwe Caroline, Touboul David, Roberts Cynthia J, Mahmoud Ashraf M, Kérautret Julien, Fournier Pierre, Malecaze François, Colin Joseph

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the biomechanical and morphological changes in keratoconic corneas after Intacs intrastromal corneal ring segment (ICRS) implantation.

SETTING

Department of Ophthalmology, National Reference Center for Keratoconus, Bordeaux University, Bordeaux, France.

METHODS

Keratoconic eyes were retrospectively analyzed after ICRS implantation; preoperative and 6-month postoperative evaluation were done using the Ocular Response Analyzer (ORA) and the Orbscan II topographer. Biomechanical parameters included corneal hysteresis (CH), the corneal resistance factor (CRF), and other parameters extracted from the signal curves. Morphological parameters included simulated keratometry and the cone location magnitude index from the axial map (aCLMI) and tangential map (tCLMI). Parameters were extracted using software designed to read and process topographic maps.

RESULTS

There were no significant differences between preoperatively and postoperatively in mean CH (7.7 mm Hg +/- 1.4 [SD] versus 7.4 +/- 1.4 mm Hg) or mean CRF (6.6 +/- 1.8 mm Hg versus 6.1 +/- 1.4 mm Hg). Only 2 ORA signal parameters were significantly different. Topographic parameters with significant decreases were minimum central keratometry (K) (mean change -5.8 +/- 2.9 diopters [D]) (P<.001), minimum central K (mean change -5.8 +/- 2.3 D) (P<.001), mean aCLMI (9.6 +/- 2.7 preoperatively versus 7.7 +/- 2.5 postoperatively) (P<.009), and mean tCLMI (18.9 +/- 2.8 versus 12.9 +/- 4.4) (P<.002). The only significant correlation between biomechanical and topographic parameters was postoperative ORA infrared signal peak 1 and postoperative aCLMI.

CONCLUSIONS

Intrastromal corneal ring implantation significantly decreased corneal curvature, with preoperative values predicting magnitude of change. However, it did not alter the viscoelastic biomechanical parameters of CH and CRF.

Для цитирования: Dauwe Caroline, Touboul David, Roberts Cynthia J, Mahmoud Ashraf M, Kérautret Julien, Fournier Pierre, Malecaze François, Colin Joseph. Biomechanical and morphological corneal response to placement of intrastromal corneal ring segments for keratoconus.. Journal of cataract and refractive surgery. 2009. DOI: 10.1016/j.jcrs.2009.05.033.