Comparison of a Novel Thin-Walled 25-gauge Needle Push Button Blood Collection Set to a Standard 23-gauge Needle in a Cancer Patient Population.

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kirchhoff Daniel, Baser Raymond, Kwong David, Ramanathan Lakshmi, McCash Samuel I

Аннотация

BACKGROUND: Phlebotomy can be an uncomfortable and even painful experience that increases in likelihood in patients who require frequent blood draws, such as those with cancer. The 25-gauge BD Vacutainer® UltraTouch™ Push Button Blood Collection Set has a smaller outer bore needle size and unique bevel configuration, which in theory should reduce pain associated with phlebotomy. Smaller needles typically cause less pain, however, they have a higher incidence of compromising the specimen integrity. Innovative engineering allows this particular needle to maintain a larger inner bore size similar to a standard 23-gauge needle. This study compares pain and specimen integrity between this novel device and a standard 23-gauge needle.

METHODS: Two cohorts of 190 patients each had phlebotomy performed with either needle. Pain experienced was assessed by patient surveys, rating pain from 0 (low) to 10 (high). Sample integrity was assessed by determining the hemolysis index (Abbott Architect c8000).

RESULTS: No statistically significant difference in median pain scores was observed between the 2 cohorts, P-value: 0.068. The 23-gauge cohort more frequently reported 3+ pain than the 25-gauge cohort, 14/190 vs 5/190. Pain scoring 1 and 2 was more frequent in the 25-gauge cohort, 84/190 vs 54/190. Pain scores of 0 were more frequent in the 23-gauge cohort, 122/190 vs 101/190. This stratification is statistically significant, P-value: 0.003. The 25-gauge needle showed no increase in hemolysis rates, P-value: 0.5.

CONCLUSIONS: Sample integrity was identical between both needles. The 25-gauge needle resulted in less 3+ pain, while pain scoring 1 and 2 was more frequent in the 25-gauge cohort.