DOI 10.1093/jalm/jfac129

Медицинская библиотека · The journal of applied laboratory medicine · 2023

Comparison of a Novel Thin-Walled 25-gauge Needle Push Button Blood Collection Set to a Standard 23-gauge Needle in a Cancer Patient Population.

Kirchhoff Daniel, Baser Raymond, Kwong David, Ramanathan Lakshmi, McCash Samuel I

СтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

BACKGROUND

Phlebotomy can be an uncomfortable and even painful experience that increases in likelihood in patients who require frequent blood draws, such as those with cancer. The 25-gauge BD Vacutainer® UltraTouch™ Push Button Blood Collection Set has a smaller outer bore needle size and unique bevel configuration, which in theory should reduce pain associated with phlebotomy. Smaller needles typically cause less pain, however, they have a higher incidence of compromising the specimen integrity. Innovative engineering allows this particular needle to maintain a larger inner bore size similar to a standard 23-gauge needle. This study compares pain and specimen integrity between this novel device and a standard 23-gauge needle.

METHODS

Two cohorts of 190 patients each had phlebotomy performed with either needle. Pain experienced was assessed by patient surveys, rating pain from 0 (low) to 10 (high). Sample integrity was assessed by determining the hemolysis index (Abbott Architect c8000).

RESULTS

No statistically significant difference in median pain scores was observed between the 2 cohorts, P-value: 0.068. The 23-gauge cohort more frequently reported 3+ pain than the 25-gauge cohort, 14/190 vs 5/190. Pain scoring 1 and 2 was more frequent in the 25-gauge cohort, 84/190 vs 54/190. Pain scores of 0 were more frequent in the 23-gauge cohort, 122/190 vs 101/190. This stratification is statistically significant, P-value: 0.003. The 25-gauge needle showed no increase in hemolysis rates, P-value: 0.5.

CONCLUSIONS

Sample integrity was identical between both needles. The 25-gauge needle resulted in less 3+ pain, while pain scoring 1 and 2 was more frequent in the 25-gauge cohort.

Для цитирования: Kirchhoff Daniel, Baser Raymond, Kwong David, Ramanathan Lakshmi, McCash Samuel I. Comparison of a Novel Thin-Walled 25-gauge Needle Push Button Blood Collection Set to a Standard 23-gauge Needle in a Cancer Patient Population.. The journal of applied laboratory medicine. 2023. DOI: 10.1093/jalm/jfac129.