The Journal of craniofacial surgery · 2020
Application of Self-Made Connection Device in Intractable Intramedullary Device.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lei Pengfei, Cui Juncheng, Wang Wu, Liu Hao, Luo Zhenhua, Yang Xucheng
Аннотация
BACKGROUND: Intramedullary fixation a standard surgical technique for long bone meta/diaphyseal fractures. There were many difficulties in removal of the intractable intramedullary device. The authors reported a new technique to remove the intractable intramedullary nail by using a self-made connecting device.
METHODS: The subject underwent removal of the intramedullary nail using a self-made connecting device, the core components of which were a caudal connecting rod and a sliding hammer in the common intramedullary nail removal device, and the auxiliary device was mainly a clinically commonly used Kirschner wire (K-wire; diameter 1.5-2.5 mm). In technical procedure, the key point was the connection between the k-wire and the intramedullary device, according to the specific conditions of the intramedullary device.
RESULTS: From 2012 to 2017, a total of 10 cases of intractable intramedullary devices were taken out using this self-made connection device, including 7 cases of tibial intramedullary nails, 1 case of femoral nail, and 1 case of tibial elastic nail. The technique provided satisfactory results, no infection or re-fracture occurred after the.
CONCLUSION: The self-made connecting device may provide new technique for more surgeons in the face of intractable intramedullary device.
METHODS: The subject underwent removal of the intramedullary nail using a self-made connecting device, the core components of which were a caudal connecting rod and a sliding hammer in the common intramedullary nail removal device, and the auxiliary device was mainly a clinically commonly used Kirschner wire (K-wire; diameter 1.5-2.5 mm). In technical procedure, the key point was the connection between the k-wire and the intramedullary device, according to the specific conditions of the intramedullary device.
RESULTS: From 2012 to 2017, a total of 10 cases of intractable intramedullary devices were taken out using this self-made connection device, including 7 cases of tibial intramedullary nails, 1 case of femoral nail, and 1 case of tibial elastic nail. The technique provided satisfactory results, no infection or re-fracture occurred after the.
CONCLUSION: The self-made connecting device may provide new technique for more surgeons in the face of intractable intramedullary device.